Appendix B to Part 136 - Definition and Procedure for the Determination of the Method Detection Limit - Revision 2
40:25.0.1.1.1.0.1.8.2 : Appendix B
Appendix B to Part 136 - Definition and Procedure for the
Determination of the Method Detection Limit - Revision 2 Definition
The method detection limit (MDL) is defined as the minimum
measured concentration of a substance that can be reported with 99%
confidence that the measured concentration is distinguishable from
method blank results.
I. Scope and Application
(1) The MDL procedure is designed to be a straightforward
technique for estimation of the detection limit for a broad variety
of physical and chemical methods. The procedure requires a
complete, specific, and well-defined analytical method. It is
essential that all sample processing steps used by the laboratory
be included in the determination of the method detection limit.
(2) The MDL procedure is not applicable to methods that
do not produce results with a continuous distribution, such as, but
not limited to, methods for whole effluent toxicity,
presence/absence methods, and microbiological methods that involve
counting colonies. The MDL procedure also is not applicable
to measurements such as, but not limited to, biochemical oxygen
demand, color, pH, specific conductance, many titration methods,
and any method where low-level spiked samples cannot be prepared.
Except as described in the addendum, for the purposes of this
procedure, “spiked samples” are prepared from a clean reference
matrix, such as reagent water, spiked with a known and consistent
quantity of the analyte. MDL determinations using spiked samples
may not be appropriate for all gravimetric methods (e.g., residue
or total suspended solids), but an MDL based on method blanks can
be determined in such instances.
II. Procedure
(1) Estimate the initial MDL using one or more of the
following:
(a) The mean determined concentration plus three times the
standard deviation of a set of method blanks.
(b) The concentration value that corresponds to an instrument
signal-to-noise ratio in the range of 3 to 5.
(c) The concentration equivalent to three times the standard
deviation of replicate instrumental measurements of spiked
blanks.
(d) That region of the calibration where there is a significant
change in sensitivity, i.e., a break in the slope of the
calibration.
(e) Instrumental limitations.
(f) Previously determined MDL.
Note:
It is recognized that the experience of the analyst is important
to this process. However, the analyst should include some or all of
the above considerations in the initial estimate of the MDL.
(2) Determine the initial MDL.
Note:
The Initial MDL is used when the laboratory does not have
adequate data to perform the Ongoing Annual Verification specified
in Section (4), typically when a new method is implemented or if a
method was rarely used in the last 24 months.
(a) Select a spiking level, typically 2 - 10 times the estimated
MDL in Section 1. Spiking levels in excess of 10 times the
estimated detection limit may be required for analytes with very
poor recovery (e.g., for an analyte with 10% recovery, spiked at
100 micrograms/L, with mean recovery of 10 micrograms/L; the
calculated MDL may be around 3 micrograms/L. Therefore, in this
example, the spiking level would be 33 times the MDL, but spiking
lower may result in no recovery at all).
(b) Process a minimum of seven spiked samples and seven method
blank samples through all steps of the method. The samples used for
the MDL must be prepared in at least three batches on three
separate calendar dates and analyzed on three separate calendar
dates. (Preparation and analysis may be on the same day.) Existing
data may be used, if compliant with the requirements for at least
three batches, and generated within the last twenty four months.
The most recent available data for method blanks and spiked samples
must be used. Statistical outlier removal procedures should not be
used to remove data for the initial MDL determination, since the
total number of observations is small and the purpose of the MDL
procedure is to capture routine method variability. However,
documented instances of gross failures (e.g., instrument
malfunctions, mislabeled samples, cracked vials) may be excluded
from the calculations, provided that at least seven spiked samples
and seven method blanks are available. (The rationale for removal
of specific outliers must be documented and maintained on file with
the results of the MDL determination.)
(i) If there are multiple instruments that will be assigned the
same MDL, then the sample analyses must be distributed across all
of the instruments.
(ii) A minimum of two spiked samples and two method blank
samples prepared and analyzed on different calendar dates is
required for each instrument. Each analytical batch may contain one
spiked sample and one method blank sample run together. A spiked
sample and a method blank sample may be analyzed in the same batch,
but are not required to be.
(iii) The same prepared extract may be analyzed on multiple
instruments so long as the minimum requirement of seven
preparations in at least three separate batches is maintained.
(c) Evaluate the spiking level: If any result for any individual
analyte from the spiked samples does not meet the method
qualitative identification criteria or does not provide a numerical
result greater than zero, then repeat the spiked samples at a
higher concentration. (Qualitative identification criteria are a
set of rules or guidelines for establishing the identification or
presence of an analyte using a measurement system. Qualitative
identification does not ensure that quantitative results for the
analyte can be obtained.)
(d) Make all computations as specified in the analytical method
and express the final results in the method-specified reporting
units.
(i) Calculate the sample standard deviation (S) of the replicate
spiked sample measurements and the sample standard deviation of the
replicate method blank measurements from all instruments to which
the MDL will be applied.
(ii) Compute the MDLs (the MDL based on spiked samples) as
follows:
MDLS =
t(n −1, 1−α = 0.99)
Ss Where: MDLs = the
method detection limit based on spiked samples
t(n-1, 1−α =
0.99) = the Student's t-value appropriate for a single-tailed 99th
percentile t statistic and a standard deviation estimate with n-1
degrees of freedom. See Addendum Table 1. Ss = sample standard
deviation of the replicate spiked sample analyses.
(iii) Compute the MDLb (the MDL based on method blanks) as
follows:
(A) If none of the method blanks give numerical results for an
individual analyte, the MDLb does not apply. A numerical result
includes both positive and negative results, including results
below the current MDL, but not results of “ND” (not detected)
commonly observed when a peak is not present in chromatographic
analysis.
(B) If some (but not all) of the method blanks for an individual
analyte give numerical results, set the MDLb equal to the highest
method blank result. If more than 100 method blanks are available,
set MDLb to the level that is no less than the 99th percentile of
the method blank results. For “n” method blanks where n ≥ 100, sort
the method blanks in rank order. The (n * 0.99) ranked method blank
result (round to the nearest whole number) is the MDLb. For
example, to find MDLb from a set of 164 method blanks where the
highest ranked method blank results are . . . 1.5, 1.7, 1.9, 5.0,
and 10, then 164 × 0.99 = 162.36 which rounds to the 162nd method
blank result. Therefore, MDLb is 1.9 for n = 164 (10 is the 164th
result, 5.0 is the 163rd result, and 1.9 is the 162nd result).
Alternatively, you may use spreadsheet algorithms to calculate the
99th percentile to interpolate between the ranks more
precisely.
(C) If all of the method blanks for an individual analyte give
numerical results, then calculate the MDLb as:
MDLb = X +
tn−1,1−α = (0.99)
Sb Where: MDLb =
the MDL based on method blanks X = mean of the method blank results
(use zero in place of the mean if the mean is negative)
t(n−1, 1α = 0.99) = the Student's t-value appropriate for
the single-tailed 99th percentile t statistic and a standard
deviation estimate with n−1 degrees of freedom. See Addendum Table
1. Sb = sample standard deviation of the replicate method blank
sample analyses. Note:
If 100 or more method blanks are available, as an option, MDLb
may be set to the concentration that is greater than or equal to
the 99th percentile of the method blank results, as described in
Section (2)(d)(iii)(B).
(e) Select the greater of MDLs or MDLb as the initial MDL.
(3) Ongoing Data Collection.
(a) During any quarter in which samples are being analyzed,
prepare and analyze a minimum of two spiked samples on each
instrument, in separate batches, using the same spiking
concentration used in Section 2. If any analytes are repeatedly not
detected in the quarterly spiked sample analyses, or do not meet
the qualitative identification criteria of the method (see section
2(c) of this procedure), then this is an indication that the
spiking level is not high enough and should be adjusted upward.
Note that it is not necessary to analyze additional method blanks
together with the spiked samples, the method blank population
should include all of the routine method blanks analyzed with each
batch during the course of sample analysis.
(b) Ensure that at least seven spiked samples and seven method
blanks are completed for the annual verification. If only one
instrument is in use, a minimum of seven spikes are still required,
but they may be drawn from the last two years of data
collection.
(c) At least once per year, re-evaluate the spiking level.
(i) If more than 5% of the spiked samples do not return positive
numerical results that meet all method qualitative identification
criteria, then the spiking level must be increased and the initial
MDL re-determined following the procedure in section 2.
(ii) [Reserved]
(d) If the method is altered in a way that can be reasonably
expected to change its sensitivity, then re-determine the initial
MDL according to section 2, and the restart the ongoing data
collection.
(e) If a new instrument is added to a group of instruments whose
data are being pooled to create a single MDL, analyze a minimum of
two spiked replicates and two method blank replicates on the new
instrument. If both method blank results are below the existing
MDL, then the existing MDLb is validated. Combine the new spiked
sample results to the existing spiked sample results and
recalculate the MDLs as in Section 4. If the recalculated MDLs does
not vary by more than the factor specified in section 4(f) of this
procedure, then the existing MDLs is validated. If either of these
two conditions is not met, then calculate a new MDL following the
instructions in section 2.
(4) Ongoing Annual Verification.
(a) At least once every thirteen months, re-calculate MDLs and
MDLb from the collected spiked samples and method blank results
using the equations in section 2.
(b) Include data generated within the last twenty four months,
but only data with the same spiking level. Only documented
instances of gross failures (e.g., instrument malfunctions,
mislabeled samples, cracked vials) may be excluded from the
calculations. (The rationale for removal of specific outliers must
be documented and maintained on file with the results of the MDL
determination.) If the laboratory believes the sensitivity of the
method has changed significantly, then the most recent data
available may be used, maintaining compliance with the requirement
for at least seven replicates in three separate batches on three
separate days (see section 2b).
(c) Include the initial MDL spiked samples, if the data were
generated within twenty four months.
(d) Only use data associated with acceptable calibrations and
batch QC. Include all routine data, with the exception of batches
that are rejected and the associated samples reanalyzed. If the
method has been altered in a way that can be reasonably expected to
change its sensitivity, then use only data collected after the
change.
(e) Ideally, use all method blank results from the last 24
months for the MDLb calculation. The laboratory has the option to
use only the last six months of method blank data or the fifty most
recent method blanks, whichever criteria yields the greater number
of method blanks.
(f) The verified MDL is the greater of the MDLs or MDLb. If the
verified MDL is within 0.5 to 2.0 times the existing MDL, and fewer
than 3% of the method blank results (for the individual analyte)
have numerical results above the existing MDL, then the existing
MDL may optionally be left unchanged. Otherwise, adjust the MDL to
the new verification MDL. (The range of 0.5 to 2.0 approximates the
95th percentile confidence interval for the initial MDL
determination with six degrees of freedom.)
Addendum to Section II: Determination of the MDL for a Specific
Matrix
The MDL may be determined in a specific sample matrix as well as
in reagent water.
(1) Analyze the sample matrix to determine the native
(background) concentration of the analyte(s) of interest.
(2) If the response for the native concentration is at a
signal-to-noise ratio of approximately 5-20, determine the
matrix-specific MDL according to Section 2 but without spiking
additional analyte.
(3) Calculate MDLb using the method blanks, not the sample
matrix.
(4) If the signal-to-noise ratio is less than 5, then the
analyte(s) should be spiked into the sample matrix to obtain a
concentration that will give results with a signal-to-noise ratio
of approximately 10-20.
(5) If the analytes(s) of interest have signal-to-noise ratio(s)
greater than approximately 20, then the resulting MDL is likely to
be biased high.
Table 1 - Single-Tailed 99th Percentile t
Statistic
Number of replicates |
Degrees of freedom
(n−1) |
t (n−1, 0.99) |
7 |
6 |
3.143 |
8 |
7 |
2.998 |
9 |
8 |
2.896 |
10 |
9 |
2.821 |
11 |
10 |
2.764 |
16 |
15 |
2.602 |
21 |
20 |
2.528 |
26 |
25 |
2.485 |
31 |
30 |
2.457 |
32 |
31 |
2.453 |
48 |
47 |
2.408 |
50 |
49 |
2.405 |
61 |
60 |
2.390 |
64 |
63 |
2.387 |
80 |
79 |
2.374 |
96 |
95 |
2.366 |
100 |
99 |
2.365 |
III. Documentation
The analytical method used must be specifically identified by
number or title and the MDL for each analyte expressed in the
appropriate method reporting units. Data and calculations used to
establish the MDL must be able to be reconstructed upon request.
The sample matrix used to determine the MDL must also be identified
with MDL value. Document the mean spiked and recovered analyte
levels with the MDL. The rationale for removal of outlier results,
if any, must be documented and maintained on file with the results
of the MDL determination.
[82 FR 40939, Aug. 28, 2017]