Title 40
SECTION 1090.1335
1090.1335 Collecting, preparing, and testing samples.
§ 1090.1335 Collecting, preparing, and testing samples.(a) General provisions. Use good laboratory practice to collect samples to represent the batch you are testing. For example, take steps to ensure that a batch is always well mixed before sampling. Also, always take steps to prevent sample contamination, such as completely flushing sampling taps and piping and pre-rinsing sample containers with the product being sampled. Follow the procedures in paragraph (b) of this section for manual sampling. Follow the procedures paragraph (c) of this section for automatic sampling. Additional requirements for measuring RVP are specified in paragraph (d) of this section. A description of how to determine compliance based on single or multiple tests on single or multiple samples is specified in paragraph (e) of this section.
(b) Manual sampling. Perform manual sampling using one of the methods specified in ASTM D4057 (incorporated by reference in § 1090.95) to demonstrate compliance with standards as follows:
(1) Collect a “running” or “all-levels” sample from the top of the tank. Drawing a sample from a standpipe is acceptable only if it is slotted or perforated to ensure that the drawn sample properly represents the whole batch of fuel.
(2)(i) Use tap sampling or spot sampling to collect upper, middle, and lower samples if a running or all-levels sample is impractical for a given storage configuration. Collect samples that most closely match the recommendations in Table 5 of ASTM D4057. Adjust spot sampling for partially filled tanks as shown in Table 1 or Table 5 of ASTM D4057, as applicable.
(ii) Spot sampling must not be used for certification testing unless the tank contains less than 10 feet of product.
(3) If the procedures in paragraphs (b)(1) and (2) of this section are impractical for a given storage configuration, you may use alternative sampling procedures as specified in ASTM D4057. This applies primarily for sampling with trucks, railcars, retail stations, and other downstream locations.
(4) Test results with manual sampling are valid only after you demonstrate homogeneity as specified in § 1090.1337.
(5) Except as specified for marine vessels in § 1090.1605, you must not do certification testing with a composite sample from manual sampling.
(c) Automatic sampling. (1) For in-line blending waivers under § 1090.1315, follow all specifications for automatic sampling as specified in EPA's approval letter instead of or in addition to the specifications in paragraph (c)(2) of this section. Automatic sampling is also appropriate for a configuration involving a pipeline filling a tank that will be certified as compliant before it leaves the fuel manufacturing facility gate.
(2) Perform automatic sampling as specified in ASTM D4177 (incorporated by reference in § 1090.95), with the following additional specifications:
(i) Configure the system to ensure a well-mixed stream at the sampling point. Align the start and end of sampling with the start and end of creating the batch.
(ii) The default sampling frequency must follow the recommended approach of at least 9,604 samples to represent a batch. Less frequent sampling is acceptable as long as the interval between samples does not exceed 20 seconds throughout the batch.
(iii) Collect three samples for individual measurements in addition to the composite sample. Draw head, middle, and tail samples after flowing 15, 50, and 85 percent of the estimated batch volume, respectively.
(iv) EPA may approve a different sampling strategy under an approved in-line blending waiver under § 1090.1315 if it is appropriate for a given facility or for a small-volume batch.
(d) Sampling provisions related to measuring RVP of summer gasoline. The following additional provisions apply for preparing samples to measure RVP of summer gasoline:
(1) Meet the additional specifications for manual and automatic sampling in ASTM D5842 (incorporated by reference in § 1090.95).
(2) If you measure other fuel parameters for a given sample in addition to RVP testing, always measure RVP first.
(e) Testing to demonstrate compliance with standards. (1) Perform testing as specified in this subpart.
(2) For parameters subject to per-gallon standards, report the highest measured value (or the lowest measured value for testing related to cetane index or other parameters that are subject to a standard representing a minimum value). This applies for repeat tests on a given sample and for testing multiple samples (including head, middle, and tail samples from automatic sampling). A batch is noncompliant if any tested sample does not meet all applicable per-gallon standards.
(3) In the case of automatic sampling for parameters subject to average standards, report the result from the composite sample to represent the batch for demonstrating compliance with the average standard. For any repeat testing with the composite sample, calculate the arithmetic average from all tests to represent the batch.
(4) In the case of manual sampling for parameters subject to average standards, determine the value representing the batch as follows:
(i) For testing with only a single sample, report that value to represent the batch. If there are repeat tests with that sample, report the arithmetic average from all tests to represent the sample.
(ii) For testing with more than one sample, report the arithmetic average from all tested samples to represent the batch. If there are repeat tests for any sample, calculate the arithmetic average of those repeat tests to determine a single value to represent that sample before calculating the average value to represent the batch.