1065.790 Mass standards.
§ 1065.790 Mass standards.(a) PM balance calibration weights. Use PM balance calibration weights that are certified as NIST-traceable within 0.1% uncertainty. Calibration weights may be certified by any calibration lab that maintains NIST-traceability. Make sure your highest calibration weight has no greater than ten times the mass of an unused PM-sample medium.
(b) Dynamometer calibration weights. [Reserved]
[70 FR 40516, July 13, 2005, as amended at 76 FR 57467, Sept. 15, 2011]