Appendix E to Subpart A of Part 82 - Article 5 Parties
40:21.0.1.1.1.1.1.24.5 : Appendix E
Appendix E to Subpart A of Part 82 - Article 5 Parties
Parties operating under Article 5 of the Montreal Protocol as of
March 26, 2014 are listed below. An updated list can be located at:
http://ozone.unep.org/new_site/en/parties_under_article5_para1.php.
Afghanistan, Albania, Algeria, Angola, Antigua & Barbuda,
Argentina, Armenia, Bahamas, Bahrain, Bangladesh, Barbados, Belize,
Benin, Bhutan, Bolivia (Plurinational State of), Bosnia and
Herzegovina, Botswana, Brazil, Brunei Darussalam, Burkina Faso,
Burundi, Cambodia, Cameroon, Cape Verde, Central African Republic,
Chad, Chile, China, Colombia, Comoros, Congo, Congo (Democratic
Republic of), Cook Islands, Cost Rica, Côte d'Ivoire, Cuba,
Djibouti, Dominica, Dominican Republic, Ecuador, Egypt, El
Salvador, Equatorial Guinea, Eritrea, Ethiopia, Fiji, Gabon,
Gambia, Georgia, Ghana, Grenada, Guatemala, Guinea, Guinea Bissau,
Guyana, Haiti, Honduras, India, Indonesia, Iran (Islamic Republic
of), Iraq, Jamaica, Jordan, Kenya, Kiribati, Korea (Democratic
People's Republic of), Korea (Republic of), Kuwait, Kyrgyzstan, Lao
(People's Democratic Republic), Lebanon, Lesotho, Liberia, Libya,
Madagascar, Malawi, Malaysia, Maldives, Mali, Marshall Islands
Mauritania, Mauritius, Mexico, Micronesia (Federal States of),
Moldova (Republic of), Mongolia, Montenegro, Morocco, Mozambique,
Myanmar, Namibia, Nauru, Nepal, Nicaragua, Niger, Nigeria, Niue,
Oman, Pakistan, Palau, Panama, Papua New Guinea, Paraguay, Peru,
Philippines, Qatar, Rwanda, Saint Kitts and Nevis, Saint Lucia,
Saint Vincent & the Grenadines, Samoa, Sao Tome and Principe, Saudi
Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore,
Solomon Islands, Somalia, South Africa, South Sudan*, Sri Lanka,
Sudan, Suriname, Swaziland, Syrian Arab Republic, Tanzania (United
Republic of), Thailand, The Former Yugoslav Republic of Macedonia,
Timor-Leste, Togo, Tonga, Trinidad and Tobago, Tunisia, Turkey,
Turkmenistan, Tuvalu, Uganda, United Arab Emirates, Uruguay,
Vanuatu, Venezuela (Bolivarian Republic of), Viet Nam, Yemen,
Zambia, Zimbabwe.
* temporarily categorized as Article 5 pending submission of ODS
consumption data [79 FR 16687, Mar. 26, 2014]
Appendix E to Subpart B of Part 82 - The Standard for Automotive Refrigerant Recycling Equipment Intended for Use With Both CFC-12 and HFC-134a
40:21.0.1.1.1.2.1.8.17 : Appendix E
Appendix E to Subpart B of Part 82 - The Standard for Automotive
Refrigerant Recycling Equipment Intended for Use With Both CFC-12
and HFC-134a
SAE J2211, Recommended Service Procedure for the Containment of
HFC-134a, as set forth under Appendix C of this subpart, and SAE
J1989, Recommended Service Procedure for the Containment of CFC-12,
as set forth under Appendix A of this subpart, also apply to this
Appendix E of this subpart.
SAE J1770, issued December, 1995.
Automotive Refrigerant Recycle Equipment Intended for Use With Both
CFC-12 and HFC-134a Foreword
The purpose of this standard is to establish specific minimum
equipment requirements for automotive refrigerant recycling
equipment intended for use with both CFC-12 and HFC-134a in a
common refrigerant circuit. Establishing such specifications will
assure that this equipment does not cross contaminate refrigerant
above specified limits when used under normal operating
conditions.
1. Scope
The purpose of this standard is to establish the specific
minimum equipment intended for use with both CFC-12 and HFC-134a in
a common refrigerant circuit that has been directly removed from,
and is intended for reuse in, mobile air-conditioning (A/C)
systems. This standard does not apply to equipment used for CFC-12
and HFC-134a having a common enclosure with separate circuits for
each refrigerant.
2. References
2.1 Applicable Documents - The following publications form a
part of this specification to the extent specified. The latest
issue of SAE publications shall apply.
2.1.1 SAE Publications - Available from SAE, 400 Commonwealth
Drive, Warrendale, PA 15096-0001.
SAE J2099 - Standard of Purity for Recycled HFC-134a for Use in
Mobile Air-Conditioning Systems SAE 1991 - Standard of Purity for
Use in Mobile Air-Conditioning Systems SAE J2196 - Service Hoses
for Automotive Air-Conditioning SAE J2197 - Service Hose Fittings
for Automotive Air-Conditioning SAE J2210 - HFC-134a (R-134a)
Recycling Equipment for Mobile A/C Systems SAE J1990 - Extraction
and Recycling Equipment for Mobile A/C Systems
2.1.2 Compressed Gas Association (CGA) Publications - Available
from CGA, 1235 Jefferson Davis Highway, Arlington, VA 22202.
CGA Pamphlet S-1.1 - Pressure Relief Device Standard
Part 1 - Cylinders for Compressed Gases
2.1.3 DOT Publications - Available from the Superintendent of
Documents, U.S. Government Printing Office, Washington, D.C.
20402
2.1.4 UL Publications - Available from Underwriters
Laboratories, 333 Pfingsten Road, Northbrook, IL 60062-2096.
UL 1769 - Cylinder Valves UL 1963 - Refrigerant Recovery/Recycling
Equipment 3. Specification and General Description
3.1 The equipment shall be suitable for use in an automotive
service garage environment and be capable of continuous operation
in ambients from 10 to 49 °C.
3.2 The equipment must be certified that it meets this
specification by Underwriters Laboratories Inc. (UL), or by an
equivalent Nationally Recognized Testing Laboratory (NRTL).
3.3 The equipment shall have a label which states “Design
Certified by (Certifying Agent) to meet SAE J1770 for recycling
CFC-12 and HFC-134a using common refrigerant circuits”, in
bold-type letters a minimum of 3 mm in height.
4. Equipment Requirements
4.1 General
4.1.1 The equipment shall be capable of preventing cross
contamination to the level required by Section 9.2.1.G before an
operation involving a different refrigerant can begin. The
equipment must prevent initiation of the recovery operation if the
equipment is not set up properly.
4.1.2 If an operator action is required to clear the unit prior
to reconnecting for a different refrigerant, the equipment shall be
provided with a means which indicates which refrigerant was last
processed.
4.1.3 Means shall be provided to prevent recovery from both an
CFC-12 and HFC-134a mobile air conditioning system
concurrently.
4.1.4 Transfer of recycled refrigerant - Recycled refrigerant
for recharging and transfer shall be taken from the liquid phase
only.
4.2 Seat Leakage Test
4.2.1 Valves, including electrically operated solenoid valves,
that are used to isolate CFC-12 and HFC-134a refrigerant circuits,
shall have a seat leakage rate not exceeding 15 g/yr ( 1/2 oz/yr)
before and after 100,000 cycles of operation. This Endurance Test
shall be conducted with HFC-134a at maximum operating pressure as
determined by sections 8.1 and 8.2. The Seat Leakage Test shall be
performed at 1.5 times this pressure at an ambient of 24 °C.
4.3 Interlocks
4.3.1 Electrical interlock devices used to prevent cross
contamination of refrigerant shall be operated for 100,000 cycles
and there shall be no failure that would permit cross contamination
of refrigerant. Solid state inter lock devices shall comply with
the Transient Overvoltage Test and the Fast Transient (Electric
Noise) Test contained in the Standard for Tests for Safety Related
Controls Employing Solid-State Devices, UL 991.
4.4 Noncondensable Gases
4.4.1 The equipment shall either automatically purge
noncondensables (NCGs) if the acceptable level is exceeded or
incorporate a device that indicates to the operator the NCG level
has been exceeded. A pressure gauge used to indicate an NCG level
shall be readable in 1 psig increments. NCG removal must be part of
the normal operation of the equipment and instructions must be
provided to enable the task to be accomplished within 30
minutes.
4.4.2 Refrigerant loss from noncondensable gas purging, oil
removal, and refrigerant clearing shall not exceed more than 5
percent by weight of the total amount of refrigerant through the
equipment as detailed in Sections 8.1, 8.2, and 9.2.
4.5 Filter
4.5.1 A 15 micron filter, or other equivalent means, to remove
particulates of 15 micrometers spherical diameter or greater shall
be located before any manual electrically operated valves that may
cause cross contamination.
4.6 Moisture and Acid
4.6.1 The equipment shall incorporate a desiccant package that
must be replaced before saturated with moisture, and whose acid
capacity is at least 5% by weight of the dry desiccant.
4.6.2 The equipment shall be provided with a moisture detection
means that will reliably indicate when moisture in the HFC-134a
exceeds 50 ppm, or in the CFC-12 exceeds 15 ppm, and requires the
filter/drier replacement.
5. Operating Instructions
5.1 The equipment manufacturer must provide operating
instructions, including proper attainment of vehicle system vacuum
(i.e., when to stop the extraction process, and also to stop
the extraction process if it is noticed that the A/C system being
serviced has a leak), filter/desiccant replacement, and purging of
noncondensable gases (air). The instructions shall indicate that
the correct sequence of operation be followed so that the equipment
can properly remove contaminates to the acceptable level. Also to
be included are any other necessary maintenance procedures, source
information for replacement parts and repair, and safety
precautions.
5.2 The equipment must prominently display the manufacturer's
name, address, the type of refrigerant (CFC-12 and HFC-134a), a
service telephone number, and the part number for the replacement
filter/drier. Operation manuals must cover information for complete
maintenance of the equipment to assure proper operation.
6. Safety Requirements
6.1 The equipment must comply with applicable federal, state,
and local requirements on equipment related to handling CFC-12 and
HFC-134a material. Safety precautions or notices related to the
safe operation of the equipment shall be prominently displayed on
the equipment and should also state “CAUTION - SHOULD BE OPERATED
BY QUALIFIED PERSONNEL”.
6.2 HFC-134a has been shown to be nonflammable at ambient
temperature and atmospheric pressure. The following statement shall
be in the operating manual: “Caution: HFC-134a service equipment or
vehicle A/C systems should not be pressure tested or leak tested
with compressed air. Some mixtures of air and HFC-134a have been
shown to be combustible at elevated pressures (when contained in a
pipe or tank). These mixtures may be potentially dangerous, causing
injury or property damage. Additional health and safety information
may be obtained from refrigerant and lubricant manufacturers.”
7. Functional Description
7.1 General
7.1.1 The equipment must be capable of ensuring recovery of the
CFC-12 and HFC-134a from the system being serviced, by reducing the
system to a minimum of 102 mm of mercury below atmospheric pressure
(i.e., vacuum).
7.1.2 The equipment must be compatible with leak detection
material that may be present in the mobile A/C system.
7.2 Shut Off Device
7.2.1 To prevent overcharge, the equipment must be equipped to
protect the tank used to store the recycled refrigerant with a
shutoff device and a mechanical pressure relief valve.
7.3 Storage Tanks
7.3.1 Portable refillable tanks or containers shall be supplied
with this equipment and must be labeled “HFC-134a” or “CFC-12” as
appropriate, meet applicable Department of Transportation (DOT) or
NRTL's Standards and be adaptable to existing refrigerant service
and charging equipment.
7.3.2 The cylinder valve shall comply with the Standard for
Cylinder Valves, UL 1769.
7.3.3 The pressure relief device shall comply with the Pressure
Relief Device Standard Part 1 - Cylinders for Compressed Gases, CGA
Pamphlet S-1.1.
7.3.4 The tank assembly shall be marked to indicate the first
retest date, which shall be 5 years after the date of manufacture.
The marking shall indicate that retest must be performed every
subsequent 5 years. The marking shall be in letters at least 6 mm
high.
7.4 Overfill Protection
7.4.1 During operation, the equipment must provide overfill
protection to assure that during filling or transfer, the tank or
storage container cannot exceed 80% of volume at 21.1 °C of its
maximum rating as defined by DOT standards, 49 CFR 173.304 and
American Society of Mechanical Engineers.
7.5 Hoses and Connections
7.5.1 Separate inlet and outlet hoses with fittings and separate
connections shall be provided for each refrigerant circuit.
7.5.2 All flexible hoses and fittings must meet SAE J2196 (for
CFC-12) and SAE J2197 (for HFC-134a).
7.5.3 Service hoses must have shutoff devices located within 30
cm of the connection point to the system being serviced.
7.6 Lubricant Separation
7.6.1 The equipment must be able to separate the lubricant from
the removed refrigerant and accurately indicate the amount of
lubricant removed during the process, in 30 mL (1 fl oz) units.
Refrigerant dissolves in lubricant and, as a result, increases the
volume of the recovered lubricant sample. This creates the illusion
that more lubricant has been recovered that actually has been. The
equipment lubricant measuring system must take into account such
dissolved refrigerant removed from the A/C system being serviced to
prevent overcharging the vehicle system with lubricant.
(Note:
Use only new lubricant to replace the amount removed the
recycling process. Used lubricant should be discarded per
applicable federal, state and local requirements.)
7.6.2 The equipment must be provided with some means, such as a
lockout device, which will prevent initiation of the recovery
operation after switching to the other refrigerant, if the
lubricant has not been drained from the oil separator.
8. Testing
8.0 Equipment shall be tested in sequence as noted in sections
8.1, 8.2 and 9.2. The filter/drier may be replaced only as noted by
section 4.6.2.
8.1 CFC-12 Recycling Cycle
8.1.1 The maximum operating pressure of the equipment shall be
determined when recycling CFC-12 while conducting the following
tests. This pressure is needed for the Seat Leakage Test, Section
4.2.
8.1.2 The equipment must be preconditioned with 13.6 kg of the
standard contaminated CFC-12 (see section 8.1.2a) at an ambient of
21 °C before starting the test cycle. Sample amounts shall be 1.13
kg with sample amounts to be repeated every 5 minutes. The sample
method fixture, defined in Figure 1 to Appendix A, shall be
operated at 21 °C.
8.1.2a Standard contaminated CFC-12 refrigerant shall consist of
liquid CFC-12 with 100 ppm (by weight) moisture at 21 °C and 45,000
ppm (by weight) mineral oil 525 suspension viscosity nominal and
770 ppm by weight of noncondensable gases (air).
8.1.3 The high moisture contaminated sample shall consist of
CFC-12 vapor with 1000 ppm (by weight) moisture.
8.1.4 The high oil contaminated sample shall consist of CFC-12
with 200,000 ppm (by weight) mineral oil 525 suspension viscosity
nominal.
8.1.5 After preconditioning as stated in section 8.1.2, the test
cycle is started, processing the following contaminated samples
through the equipment.
A. 13.6 kg (1.13 kg per batch) of standard contaminated CFC-12. B.
1 kg of high oil contaminated CFC-12. C. 4.5 kg (1.13 kg per batch)
of standard contaminated CFC-12. D. 1 kg of high moisture
contaminated CFC-12. 8.1.6 The CFC-12 is to be cleaned to the
minimum purity level, as defined in SAE J1991, with the equipment
operating in a stable ambient of 10, 21, and 49 °C and processing
the samples as defined in section 8.1.5.
8.2 HFC-134a Recycling Cycle
8.2.1 The maximum operating pressure of the equipment shall be
determined when recycling HFC-134a while conducting the following
tests. This pressure is needed for the Seat Leakage Test, Section
4.2.
8.2.2 The equipment must be preconditioned by processing 13.6 kg
of the standard contaminated HFC-134a (see section 8.2.2a) at an
ambient of 21 °C before starting the test cycle. 1.13 kg samples
are to be processed at 5 minute intervals. The text fixture shown
in Figure 1 to Appendix A shall be operated at 21 °C.
8.2.2a The standard contaminated refrigerant shall consist of
liquid HFC-134a with 1300 ppm (by weight) moisture (equivalent to
saturation at 38°[100 °F]), 45,000 ppm (by weight) HFC-134a
compatible lubricant, and 1000 ppm (by weight) of noncondensable
gases (air).
8.2.2b The HFC-134a compatible lubricant referred to in section
8.2.2a shall be a polyalkylene glycol based synthetic lubricant or
equivalent, which shall contain no more than 1000 ppm by weight of
moisture.
8.2.3 Following the preconditioning procedure per section 8.2.2,
18.2 kg of standard contaminated HFC-134a are to be processed by
the equipment at each stable ambient temperature of 10, 21, and 49
°C.
8.2.4 The HFC-134a is to be cleaned to the purity level, as
defined in SAE J2099.
9. Refrigerant Cross Contamination Test
9.1 General
9.1.1 For test validation, the equipment is to be operated
according to the manufacturer's instruction.
9.1.2 The equipment shall clean the contaminated CFC-12
refrigerant to the minimum purity level as defined in Appendix A,
when tested in accordance with the requirements in section 8.1.
9.1.3 The equipment shall clean the contaminated HFC-134a
refrigerant to the purity level defined in Appendix C, when tested
in accordance with the requirements in section 8.2.
9.2 Test Cycle
9.2.1 The following method shall be used after the tests and
requirements in Sections 8.1 and 8.2, respectively, are completed.
Following the manufacturer's instructions, the equipment shall be
cleared of HFC-134a, prior to beginning step A. The only
refrigerant used for this is noted in steps A, C, and E of section
9.2.1. The test fixture shown in Figure 1 to Appendix A shall be
used and the test shall be conducted at 10, 21, and 49 °C
ambients.
A. A 1.13 kg standard contaminated sample of CFC-12 (see section
8.1.2a) shall be processed by the equipment. B. Follow
manufacturer's instructions to clear the equipment of CFC-12 before
processing HFC-134a. C. Process a 1.13 kg, standard contaminated
sample of HFC-134a (see section 8.2.2a) through the equipment. D.
Follow manufacturer's instructions to clear the equipment of
HFC-134a before processing CFC-12. E. Process a 1.13 kg standard
contaminated sample of CFC-12 (see section 8.1.2a) through the
equipment. F. Follow manufacturer's instructions to clear the
equipment of CFC-12. G. The amount of cross contaminated
refrigerant, as determined by gas chromatography, in samples
processed during steps C and E of section 9.2.1., shall not exceed
0.5 percent by weight. 10. Sample Analysis
10.1 General
10.1.1 The processed contaminated samples shall be analyzed
according to the following procedure.
10.2 Quantitative Determination of Moisture
10.2.1 The recycled liquid phase sample of refrigerant shall be
analyzed for moisture content via Karl Fischer coulometer titration
or an equivalent method. The Karl Fischer apparatus is an
instrument for precise determination of small amounts of water
dissolved in liquid and/or gas samples.
10.2.2 In conducting the test, a weighed sample of 30 to 130 g
is vaporized directly into the Karl Fischer anolyte. A coulometer
titration is conducted and the results are calculated and displayed
as parts per million moisture (weight).
10.3 Determination of Percent Lubricant
10.3.1 The amount of lubricant in the recycled sample of
refrigerant/lubricant is to be determined by gravimetric
analysis.
10.3.2 Following venting of noncondensable, in accordance with
the manufacturer's operating instructions, the refrigerant
container shall be shaken for 5 minutes prior to extracting samples
for test.
10.3.3 A weighed sample of 175 to 225 g of liquid
refrigerant/lubricant is allowed to evaporate at room temperature.
The percent lubricant is to be calculated from the weight of the
original sample and the residue remaining after the
evaporation.
10.4 Noncondensable Gas
10.4.1 The amount of noncondensable gas is to be determined by
gas chromatography. A sample of vaporized refrigerant liquid shall
be separated and analyzed by gas chromatography. A Propak Q column
at 130 °C and a hot wire detector may be used for analysis.
10.4.2 This test shall be conducted on liquid phase samples of
recycled refrigerant taken from a full container as defined in 7.4
within 30 minutes following the proper venting of noncondensable
gases.
10.4.3 The samples shall be shaken for at least 15 minutes prior
to testing while at a temperature of 24 °C ±2.8 °C.
10.5 Refrigerant Cross Contamination
10.5.1 The amount of cross contamination of CFC-12 in HFC-134a
or HFC-134a in CFC-12 shall not exceed 0.5 percent by weight as
determined by gas chromatography. A sample of vaporized refrigerant
liquid shall be separated and analyzed by gas chromatography. A 1%
SP-1000 on Carbopack B (60/80 mesh) column may be used for the
analysis.
[62 FR 68053, Dec. 30, 1997]
Appendix E to Subpart F of Part 82 - Test Procedure for Leaks From Containers Holding Two Pounds or Less of Refrigerant for Use in an MVAC
40:21.0.1.1.1.6.1.15.27 : Appendix E
Appendix E to Subpart F of Part 82 - Test Procedure for Leaks From
Containers Holding Two Pounds or Less of Refrigerant for Use in an
MVAC
This appendix is based on the California Air Resources Board
(CARB) standard TP-503: Test Procedure for Leaks from Small Cans
of Automotive Refrigerant, as amended on January 5, 2010; and
CARB standard BP-A1: Balance Protocol for Gravimetric
Determination of Sample Weights using a Precision Balance, as
amended January 5, 2010.
Section 1. Applicability
This test procedure is used by manufacturers of containers
holding two pounds or less of refrigerant for use in a motor
vehicle air conditioner (MVAC) to determine the leakage rate of
small containers of automotive refrigerant that are subject to the
requirements of 40 CFR part 82, subpart F. Specifically, this test
procedure will specify the equipment, procedures, and calculations
to determine if a container holding two pounds or less of
refrigerant for use in an MVAC complies with the leakage rate
specified in § 82.154(c)(2)(ii). All terms in this appendix will
follow the definitions in § 82.152 unless otherwise defined in this
appendix.
All containers holding two pounds or less of refrigerant for use
in an MVAC must comply with other applicable codes and regulations
such as local, state, or Federal safety codes and regulations.
This test procedure involves the use of materials under pressure
and operations and should only be used by or under the supervision
of those familiar and experienced in the use of such materials and
operations. Appropriate safety precautions should be observed at
all times while performing this test procedure.
Section 2. Principle and Summary of Test Procedure
This procedure is used to determine the leakage rate of
containers holding two pounds or less of refrigerant for use in an
MVAC (small cans). Testing will involve subjecting both full and
partially empty cans in both upright and inverted positions at two
temperatures: 73 °F and 130 °F.
Thirty small cans are tested under each condition for a total of
240 small cans tested. Small cans are brought to temperature
stability, weighed, then stored for 30 days under specified
conditions of temperature, orientation, and state of fill, then
re-weighed. Leakage rate (grams/year) is estimated by (weight loss
in grams) x 365/(days duration). The leakage rate is then compared
to a standard of 3.00 grams/year to determine if a given small can
complies with the leakage rate specified in § 82.154(c)(2)(ii).
Section 3. Biases and Interferences
3.1 Contaminants on the operator's hands can affect the weight
of the small can and the ability of the small can to absorb
moisture. To avoid contamination of the small can, the balance
operator should wear gloves while handling the small cans.
3.2 Weight determinations can be interfered with by moisture
condensing on the small can and by thermal currents generated by
temperature differences between the small can and the room
temperature. The small cans cool during discharge and could cause
condensation. For these reasons, small cans must be equilibrated to
balance room temperature for at least four hours before
weighing.
3.3 Variations in the temperature, pressure, and humidity of the
ambient air will cause variations in the buoyancy of the small can.
These variations should typically be less than 25 mg for a small
can. If the small can is not leaking at all, then the uncorrected
weight changes will be within the range of 0 ± 25 mg, which is
about ten percent of the 247 mg loss expected after thirty days for
a can leaking at 3 g/yr. In that case buoyancy corrections can be
omitted. If the absolute value of the uncorrected weight change
exceeds 25 mg, then all calculations must be made using weights
corrected for buoyancy based on the temperature, pressure, and
humidity of the weighing room.
3.4 Some electronic balances are sensitive to the effects of
small static charges. The small can should be placed directly on
the balance pan, ensuring metal to metal contact. If the balance
pan is not grounded, the small can and balance pan should be
statically discharged before weighing.
Section 4. Sensitivity and Range
The mass of a full small can could range from roughly 50 g to
1000 g depending on the container capacity. A top loading balance,
capable of a maximum weight measurement of not less than 1,000 g
and having a minimum readability of 0.001 g, reproducibility and
linearity of ± 0.002 g, must be used to perform mass
measurements.
Section 5. Equipment
5.1 A top loading balance that meets the requirements of Section
4 above.
5.2 A NIST traceable working standard mass for balance
calibration. A NIST traceable working standard mass for a balance
linearity check. A reference mass to serve as a “blank” small
can.
5.3 An enclosure capable of controlling the internal air
temperature from 73 °F ± 5 °F, and an enclosure capable of
controlling the internal air temperature to 130 °F ± 5 °F.
5.4 A temperature instrument capable of measuring the internal
temperature of the temperature conditioning enclosures and the
balance room with a sensitivity of ± 2 °F.
5.5 A barometric pressure instrument capable of measuring
atmospheric pressure at the location of the balance to within ±
0.02 inches of mercury.
5.6 A relative humidity measuring instrument capable of
measuring the relative humidity (RH) at the location of the balance
with a sensitivity of ± 2 percent RH.
5.7 A hose with appropriate fitting for dispensing refrigerant
from the small can to a recovery machine.
5.8 A refrigerant recovery machine to collect the discharged
refrigerant from small cans being tested.
Section 6. Calibration Procedures
6.1 Calibrations are applied to the balance and to the support
equipment such as temperature, humidity, and pressure monitoring
equipment. Procedures for calibration are not spelled out here.
General calibration principals for the support equipment and the
balance are described in Section 11, Quality Assurance/Quality
Control. Detailed calibration procedures for measurements made
using the balance are contained in Attachment A: “Balance Protocol
for Gravimetric Determination of Sample Weights using a Precision
Balance.”
Section 7. Small Can Preparation
7.1 Receive a batch of 240 small cans of one design to be
tested. These may include several SKUs from different manufacturers
if the container and valve combination are the same.
7.2 Clean small cans with Alkanox solution or equivalent and dry
with a lint free towel.
7.3 Confirm that the sample ID sticker on the small can matches
the sample ID on the chain of custody forms.
7.4 Select a reference mass similar to the weight of a full
small can. If multiple sets of similar sized small cans are being
tested, only one reference mass is needed; it can be used with all
sets. Store the reference mass in the balance area.
7.5 Evacuate the contents of one half of the small cans (120
cans) into the refrigerant recovery machine using normal DIY
dispensing procedures until each small can is approximately half
full.
7.6 Select a reference mass similar to the weight of the
half-full small can. If multiple sets of similar size small cans
are being tested, only one reference mass is needed; it can be used
with all sets. Store the reference mass in the balance area.
Section 8. Small Can Weighing
Weighing cans on the balance is done in accordance with
Attachment A to this appendix. Attachment A describes how to
conduct weight determinations including appropriate calibration and
QC data. This section, “Small Can Weighing,” describes the overall
process, not the details of how to use the balance.
Initial Weights
8.1 Put on gloves. Check the small cans for contamination.
8.2 Place the 240 small cans into a location where they can
equilibrate to balance room temperature. Record the small can test
IDs and the equilibration start time on the Small Can Test Data
Forms available on EPA's Web site in sets of thirty, one form for
each of the eight test conditions.
8.3 Let cans equilibrate for at least four hours.
8.4 Weigh the set of 240 small cans and the reference weights
using Attachment A and log the results to the Balance Weighing Log
Form available on EPA's Web site.
8.5 Transfer data from the Balance Weighing Log Form to the
Small Can Test Data Form in sets of 30, one set for each of the
eight conditions to be tested.
Thirty-Day Soak
8.6 Place each set of 30 small cans into the appropriate
orientation and temperature for soaking:
30 full small cans - 73 °F, upright 30 full small cans - 73 °F,
inverted 30 full small cans - 130 °F, upright 30 full small cans -
130 °F, inverted 30 half-full small cans - 73 °F, upright 30
half-full small cans - 73 °F, inverted 30 half-full small cans -
130 °F, upright 30 half-full small cans - 130 °F, inverted
8.7 Soak the small cans for 30 days undisturbed.
Final Weighing
8.8 Place the 240 small cans into a location where they can
equilibrate to balance room temperature.
8.9 Let the small cans equilibrate for at least four hours.
8.10 Weigh the set of 240 small cans, the reference weights, and
any additional sets of small cans using Attachment A.
8.11 Transfer data from the Balance Weighing Log Form to the
corresponding Small Can Test Data Forms.
Section 9. Calculations Corrections for Buoyancy
The calculations in this section are described in terms of
“weight.” Mass is a property of the small can, whereas weight is a
force due to the effects of buoyancy and gravity. Procedures for
correcting the effect of buoyancy are given in Attachment B of this
appendix. Ignoring buoyancy, i.e., using weight data
uncorrected for buoyancy effects, is acceptable for a thirty day
test if the absolute magnitude of the weight change is less than 25
mg. If the uncorrected weight change exceeds 25 mg for any small
can, then correct all small can weights for buoyancy using the
procedures in Attachment B before performing the calculations
described below.
Calculation of Leak Rate
The emission rate in grams/day for each small can is calculated
by subtracting the final weight from the initial weight and then
dividing the weight difference by the time difference measured in
days to the nearest hour (nearest 1/24 of a day). The emission rate
in g/day is multiplied by 365 to determine emission rate in
grams/yr. If the annual emission rate for any small can exceeds the
entire small can contents, then the annual emission rate for that
small can is adjusted to equal the entire small can contents/year
(e.g., about 350 g/yr for a 12 ounce small can). The annual
emission rate for the purpose of the test is calculated by
averaging the 240 individual adjusted annual emission rates and
rounding to two decimal places. The cans fail the test if the
adjusted annual emission rate averaged over 240 cans is greater
than 3.00 g/yr. The calculations are described below.
Loss rate for each small can
E
idaily = (W
ifinal −
W
iinitial)/(D
ifinal − D
iinitial) g/day
E
iannual = 365 × E
idaily g/year E
iadjusted =
Minimum of (E
iadjusted, C
i/year) g/yr Where,
E
i = emission rate W
ifinal = weight of can
i
after soaking (grams) W
iinitial = weight of can I before
soaking (grams) D
ifinal = date/time of final weight
measurements (days) D
iinitial = date/time of initial weight
measurements (days) C
i = original factory mass of
refrigerant in can
i
Note: Date/Times are measured in days. Microsoft Excel
stores dates and times in days, and the calculations can be made
directly in Excel. If calculations are made manually, calculate
serial days to the nearest hour for each date and time as
follows:
D = Julday + Hour/24 Where, Julday = serial day of the year: Jan 1
= 1, Jan 31 = 31, Feb 1 = 32, etc. Hour = hour of day using 24-hour
clock, 0 to 23
Calculate the average loss rate for the 240 small cans as
follows:
Emean = [Sum (Eadjusted
i),
i = 1 to 240]/240 Section
10. Recordkeeping
During small can weighing, record the small can weights and
date/times on the Balance Weighing Log Form. After each weighing
session, transfer the measured weights and date/times from the
Balance Weighing Log Form to the Small Can Test Data Form.
At the end of the test, complete the calculations described in
Section 9, Calculations, and record the results on the Small Can
Test Data Form.
Section 11. Quality Assurance/Quality Control
11.1 All temperature, pressure, and humidity instruments should
be calibrated annually against NIST traceable laboratory standards.
The main purpose of the NIST traceable calibration is to establish
the absolute accuracy of the device. The instruments should also be
checked periodically such as weekly, monthly, or quarterly against
intermediate standards or against independent instruments. For
example, a thermocouple can be checked weekly against a wall
thermometer. A barometer or pressure gauge can be checked weekly by
adjusting to sea level and comparing with local airport data. The
main purpose of the frequent checks is to verify that the device
has not failed in some way. This is especially important for
electronic devices such as a digital thermometer, but even a liquid
filled thermometer can develop a problem such as a bubble.
11.2 The balance should be serviced and calibrated annually by
an independent balance service company or agency using NIST
traceable reference masses. Servicing verifies accuracy and
linearity, and the maintenance performed helps ensure that a
malfunction does not develop.
11.3 The balance must also be calibrated and its linearity
checked with working standards before and after each weighing
session, or before and after each group of 24 small cans if more
than 24 small cans are weighed in a session. Procedures for
calibrating and using the balance, as well as recording balance
data, are described in the accompanying balance weighing protocol.
These procedures include zero checks, calibration checks, and
reference mass checks. Procedures for calculating quality control
data from those checks are described in Attachment A.
11.4 The small cans are cleaned then handled using gloves to
prevent contamination. All equilibration and soaking must be done
in a dust free area.
Section 12. Balance Protocol for Gravimetric Determination of
Sample Weights Using a Precision Balance
12.1 Scope and application
This Protocol summarizes a set of procedures and tolerances for
weighing objects in the range of 0 to 1,000 g with a resolution of
0.001 g. This protocol only addresses balance operations, it does
not address project requirements for equilibration, sample hold
time limits, sample collection etc.
12.2 Summary of method
The balance is zeroed and calibrated using procedures defined
herein. Object weight determinations are conducted along with
control object weight determinations, zero checks, calibration
checks, sensitivity checks, and replicate weightings in a defined
sequence designed to control and quantitatively characterize
precision and accuracy.
12.3 Definitions
N/A.
12.4 Interferences
Object weights can be affected by temperature and relative
humidity of their environment, air currents, static electricity,
gain and loss of water vapor, gain or loss of and loss of volatile
compounds directly from the sample or from contaminants such as
finger prints, marker ink, and adhesive tape.
Contamination, transfer of material to or from the samples, is
controlled by conducting operations inside a clean area dedicated
to the purpose and having a filtered laminar air flow where
possible; by wearing gloves while handling all samples and related
balance equipment; by using forceps to handle small objects, and by
keeping the balance and all related equipment inside the clean
area.
Air currents are controlled by conducting weighing operations
inside a closed chamber or glove box and by allowing the substrates
to reach temperature and relative humidity equilibrium. The chamber
is maintained at 40 percent relative humidity and 25 °C by a
continuous humidity and temperature control system. The temperature
and RH conditions are recorded at least once per weighing sessions.
Equilibration times for samples that are particularly sensitive to
humidity or to loss of semi-volatiles species are specified by
project requirements.
Static electric charges on the walls of the balance and the
weighed objects, including samples, controls, and calibration
weights, can significantly affect balance readings. Static is
avoided by the operator ground himself and test objects as
described in the balance manual.
12.5 Personnel health and safety
N/A
12.6 Equipment and supplies
• Filtered, temperature and humidity controlled weighing
chamber.
• Precision Balance
• Plastic forceps
• Nylon fabric gloves.
• Working calibration weights: ANSI Class 2, 1000g and 500 g
• Working sensitivity weight: 50 mg
• Reference objects: references are one or more objects that are
typical of the objects to be weighed during a project, but that are
stored permanently inside the balance glove box. Reference objects
are labeled Test1, Test2, Test3, etc.
12.7 Reagents and standard
N/A
12.8 Sample collection, preservation, and storage
N/A. See relevant project requirements and SOPs.
12.9 Quality control
Data quality is controlled by specifying frequencies and
tolerances for Zero, Calibration, Linearity, and Sensitivity
checks. If checks do not meet tolerance criteria, then samples must
be re-weighed. In addition, the procedures specify frequencies for
Control Object Checks.
Data quality is quantitatively characterized using Zero Check,
Calibration Check, and Control Check data. These data are
summarized monthly in statistics and QC charts.
12.10 Calibration and standardization
The absolute accuracy of the balance is established by
calibration against an ANSI Class 2, stainless steel working
weight: 1000.000 g ± 0.0025 g. Linearity is established checking
the midpoint against an ANSI Class 2 stainless steel working
weight: 500.000 ± 0.0012 g. Sensitivity is established using and
ANSI Class 2 stainless steel or aluminum working weight: 50 mg.
Precision is checked by periodically checking zero, calibration,
and reference object weights.
12.11 Procedure
12.11.1 Overview of Weighing Sequence
Weighing a series of substrates consists of performing the
following procedures in sequence, while observing the procedures
for handling and the procedures for reading the balance:
1. Initial Adjustment 2. Weigh eight samples 3. Zero Check 4. Weigh
eight samples 5. Zero Check 6. Weigh eight samples 7. Calibration
Check 8. Return to step 2. 9. If less than 24 cans are weighed,
perform a final Calibration Check at the end of weighing.
This sequence is interrupted and samples are reweighed if QC
check tolerances are not met. Each of these procedures along with
procedures for handling and reading the balance are described
below. The QC tolerances referred to in these procedures are listed
in Table 1.
12.11.2 Handling
1. Never touch samples, weights, balance pans, etc. with bare
hands. Wear powder free gloves to handle the weights, controls, and
samples.
12.11.3 Reading the Balance
1. Close the door. Wait for the balance stabilization light to
come on, and note the reading.
2. Watch the balance reading for 30 sec (use a clock). If the
reading has not changed by more than 0.001 g from the reading noted
in step 1, then record the reading observed at the end of the 30
sec period.
3. If the reading has drifted more than 0.001 g note the new
balance reading and go to step 2.
4. If the balance reading is flickering back and forth between
two consecutive values choose the value that is displayed more
often than the other.
5. If the balance reading is flickering equally back and forth
between two consecutive values choose the higher value.
12.11.4 Initial Adjustment
1. Empty the sample pan Close the door. Select Range 1000 g 2. Wait
for a stable reading 3. Record the reading with QC code IZC
(initial zero check) 4. Press the Tare button 5. Record the reading
in the logbook with QC code IZA (initial zero adjust) 6. Place the
1,000 g working calibration weight on the balance pan 7. Wait for a
stable reading. 8. Record the reading with QC code ICC (initial cal
check) 9. Press the Calibrate button 10. Record the reading with QC
code ICA (initial cal adjust) 11. Remove the calibration weight.
12. Wait for a stable reading. 13. Record the reading with QC code
IZC. 14. If the zero reading exceeds ± 0.002 g, go to step 4. 15.
Place the 500 g calibration weight on the balance pan 16. After a
stable reading, record the reading with QC code C500. Do not adjust
the balance. 17. Add the 0.050 g weight to 500 g weight on the
balance pan. 18. After a stable reading, record the reading with QC
code C0.05. Do not adjust the balance. 19. Weigh reference object
TEST1, record reading with QC code T1. 20. Weigh the reference
object TEST2, TEST3, etc. that is similar in weight to the samples
that you will be weighing. Record with QC code T2, T3, etc.
12.11.5 Zero Check
1. Empty the sample pan. Close the door. 2. Wait for a stable
reading 3. Record the reading with QC code ZC 4. If the ZC reading
is less than or equal to the zero adjustment tolerance shown in
Table 1, return to weighing and
do not adjust the zero. If
the ZC reading exceeded the zero adjustment tolerance, proceed with
steps 5 through 7. 5. Press the Tare button 6. Record the reading
in the logbook with QC code ZA. 7. If the ZC reading exceeded the
zero re-weigh tolerance, change the QC code recorded in step 3 from
ZC to FZC. Then enter a QC code of FZ into the QC code column of
all samples weights obtained after the last valid zero check. Re-
weigh all of those samples, recording new data in new rows of the
logbook.
12.11.6 Calibration Check
1. First, follow procedures for Zero Check. If the ZC was within
tolerance, tare the balance anyway (
i.e., follow steps 5 and
6 of the Zero Check method) 2. Place the 1,000 g working
calibration weight on the sample pan, wait for a stable reading. 3.
Record the reading with QC code C1000 4. If the C1000 reading is
less than or equal to the calibration adjustment tolerances, skip
steps 5 through 8 and proceed to step 9.
Do not adjust the
calibration. 5. If the C100 reading exceeded the calibration
adjust tolerance, press the Calibrate button. 6. Record the reading
in the logbook with QC code CA 7. Perform a Zero Check (follow the
Zero Check method) 8. If the C1000 reading exceeded the calibration
re-weigh tolerance, change the code recorded in step 3 from C1000
to FC1000. Enter FC into the QC column for all sample weights
obtained after the last valid calibration check. Re-weigh all of
those samples, recording new data in new rows of the logbook.
12.11.7 Replicate Weighing Check
1. This protocol does not include reweigh samples to obtain
replicates. The projects for which this protocol is intended
already include procedures multiple weightings of each sample.
Table 1 - QC Tolerances and Frequencies for
Balance Protocol
Reading Tolerance: |
0.001
g, stable for 30 sec. |
Adjustment Tolerances: |
Zero: |
−0.003 to +0.003 g. |
Calibration: |
999.997 to 1000.003 g. |
Controls: |
none. |
Replicates: |
none. |
Re-weigh Tolerances: |
Zero: |
−0.005 to +0.005 g. |
Calibration: |
999.995 to 1000.005 g. |
Controls: |
none. |
Replicates: |
none. |
Reference Objects: |
Test 1
- A reference object weighing about 400 g. |
Test 2
- A reference object weighing about 200 g. |
Test 3
- A reference object weighing about 700 g. |
QC
Frequencies: |
Zero Checks: |
once per 8 samples. |
Calibration
Checks: |
once per 24 samples. |
Repeat
weighings: |
none (test method includes
replicate determinations). |
Control
objects: |
once per weighing
session. |
12.12 Data analysis and calculations
For Zero Checks, let Z equal the recorded Zero Check value. For
control checks let T1, T2, etc. equal the recorded value for
control object Test 1, Test 2, etc. For Calibration Checks, let
C1000 equal C1000 reading minus 1000, M = C500 - 500, S = .C.050 -
C500 - .050. For Replicate Checks, let D equal the loss that
occurred between the first and second measurements. In summary:
T1 = T1 T2 = T2 T3 = T3 Z = ZC - 0 C = C1000 - 1000 M = C500 - 500
G = C050 - C500 - .050
Tabulate the mean and standard deviation for each of the
following: Z, C, M, G. T1, T2, T3. Depending on the number of
operators using the balance and the number of protocols in use,
analyze the data by subcategories to determine the effects of
balance operator and protocol. Each of these standard deviations,
SZ, SC, etc. is an estimate of the precision of single weight
measurement.
For Z, C, M, and G, check the mean value for statistical
difference from 0. If the means are statistically different than
zero, troubleshooting to eliminate bias may be called for. For Z,
C, M, G, T1, T2, T3, check that the standard deviations are all
comparable. If there are systematic differences, then
troubleshooting to eliminate the problem may be called for.
Note that the precision of a weight gain, involves two weight
determinations, and therefore is larger than S by a factor of
sqrt(2). On the other hand replicate weighings improves the
precision of the determinations by a factor of sqrt(N). If N = 2,
i.e., duplicates, then the factors cancel each other.
To estimate the overall uncertainty in a weight determination, a
conservative estimate might be to combine the imprecision
contributed by the zero with the imprecision contributed by the
calibration.
U = Sqrt(SZ 2 + SC 2)
The uncertainty in a weight gain from N replicates is then given
by:
Ugain = Sqrt(2) × Sqrt(SZ 2 + SC 2)/Sqrt(N)
But due to the balance adjustment and reweigh tolerances, we
expect SZ to approximately equal SC, to approximately equal SM,
etc. tolerances, so that the equation above becomes:
Ugain = 2 × S/Sqrt(N) Where S is any individual standard deviation;
or better, a pooled standard deviation.
12.13 Method performance
The data necessary to characterize the accuracy and precision of
this method are still being collected. The method is used primarily
to weigh objects before and after a period of soaking to determine
weight loss by subtraction. Given the reweigh tolerances, we expect
that the precision of weight gain determinations will be on the
order of 0.006 g at the 1-sigma level. Bias in the weight gain
determination, due to inaccuracy of the calibration weight and to
fixed non-linearity of the balance response is on the order 0.005
percent of the gain.
12.14 Pollution prevention
When discharging half the can contents during can preparation,
do not vent the contents of the small can to the atmosphere. Use an
automotive recovery machine to transfer small can contest to a
recovery cylinder.
12.15 Waste management
Dispose of the contents of the recycle cylinder through a
service that consolidates waste for shipment to EPA certified
facilities for reclaiming or destruction.
Section 13. Compensation of Weight Data for Buoyancy and Gravity
Effects
13.1 Gravity
Variations in gravity are important only when weighing objects
under different gravitational fields, i.e., at different
locations or at different heights. Since the balance procedures
calibrate the balance against a known mass (the calibration
“weight”) at the same location where sample objects are weighed,
there is no need to correct for location. Although both the sample
and the calibration weight are used at the same location, there
will be a difference in the height of the center of gravity of the
sample object (small can) and the center of gravity of the
reference mass (calibration weight). However, this difference in
height is maintained during both the initial weights and final
weights, affecting the initial and final weights by the same
amount, and affecting the scale of the weight difference by only a
few ppm. In any event, the magnitude of this correction is on the
order of 0.3 ug per kg per mm of height difference. A difference on
the order of 100 mm would thus yield a weight difference of about
0.03 mg, which is insignificant compared to our balance resolution
which is 0.001 g or 1 mg.
Based on the discussion above, no corrections for gravity are
necessary when determining weight changes in small cans.
13.2 Buoyancy
Within a weighing session, the difference in density between the
sample object and the calibration weight will cause the sample
object weight value to differ from its mass value due to buoyancy.
For a 1-liter object in air at 20 °C and at 1 atm, the buoyant
force is about 1.2 g. The volume of a 1 kg object with a density of
8 g/cm 3 (e.g., a calibration weight), is about 0.125
liters, and the buoyancy force is about 0.15 g. Variations in air
density will affect both of these values in proportion. The net
value being affected by variations in air density is thus on the
order of 1.2 − 0.15 = 1.05 g. Air density can vary up or down by 2
percent or more due to variations in barometric pressure,
temperature, and humidity. The buoyancy force will then vary up or
down by 0.02 g, or 20 mg. This is significant compared to the
weight change expected after one week for a can leaking at 3 grams
per year, which is 57 mg.
Based on the discussion above, buoyancy corrections must be
made.
Variables measured or calculated:
Vcan = volume of can (cm 3). Estimate to within 10 percent by
measuring the can dimensions or by water displacement. Error in the
can volume will cause an error in the absolute amount of the
buoyancy force, but will have only a small effect on the change in
buoyancy force from day to day. Wcan = nominal weight of a can (g),
used to calculate the nominal density of the can. ρcan = nominal
density of a small can (g/cm 3). The nominal values can be applied
to corrections for all cans. It is not necessary to calculate a
more exact density for each can. Calculate once for a full can and
once for a half full can as follows: ρcan = Wcan /Vcan T =
Temperature in balance chamber (degrees Celsius). RH = Relative
humidity in balance chamber (expressed a number between 0 and 100).
Pbaro = Barometric pressure in balance chamber (millibar). Use
actual pressure, NOT pressure adjusted to sea level. ρair = density
of air in the balance chamber (g/cm 3). Calculate using the
following approximation: ρair = 0.001*[0.348444*Pbaro−(RH/100) ×
(0.252 × T−2.0582)]/(T + 273.15) ρref = the reference density of
the calibration weight (g/cm 3). Should be 8.0 g/cm 3. Equation to
correct for buoyancy: Wcorrected = Wreading × (1 - ρair/ρref)/(1 -
ρair/ρcan) [81 FR 82392, Nov. 18, 2016]