Appendix to Part 1195 - Standards for Accessible Medical Diagnostic Equipment
36:3.0.9.1.11.0.5.2.20 :
Appendix to Part 1195 - Standards for Accessible Medical Diagnostic
Equipment Table of Contents Chapter 1: Application and
Administration M101 General M102 Definitions Chapter 2: Scoping
M201 General Chapter 3: Technical Requirements M301 Diagnostic
Equipment Used by Patients in Supine, Prone, or Side-Lying Position
M302 Diagnostic Equipment Used by Patients in Seated Position M303
Diagnostic Equipment Used by Patients Seated in a Wheelchair M304
Diagnostic Equipment Used by Patients in Standing Position M305
Supports M306 Communication M307 Operable Parts Chapter 1:
Application and Administration M101 General
M101.1 Purpose. These Standards (MDE Standards) contain
scoping and technical requirements for medical diagnostic equipment
(diagnostic equipment) to ensure accessibility to, and usability of
the diagnostic equipment by patients with disabilities. The MDE
Standards provide for independent access to, and use of, diagnostic
equipment by patients with disabilities to the maximum extent
possible.
M101.2 Application. Sections M301 through M304 shall be
applied to diagnostic equipment, based on the patient positions
that the equipment supports, during patient transfer and diagnostic
use. Sections M306 and M307 shall be applied to diagnostic
equipment where communication features or operable parts are
provided for patient use.
M101.3 Existing Diagnostic Equipment. The MDE Standards
do not address the applicability of scoping or technical
requirements to existing diagnostic equipment. Enforcing
authorities, such as the Department of Justice or the Department of
Health and Human Services, have authority over the accessibility of
existing equipment and any regulation of that equipment will be
effective only to the extent required by such enforcing
authorities.
M101.4 Equivalent Facilitation. The use of alternative
designs or technologies that result in substantially equivalent or
greater accessibility and usability than specified in the MDE
Standards is permitted.
M101.5 Dimensions. The MDE Standards are based on adult
dimensions and anthropometrics. Dimensions that are not stated as
“maximum” or “minimum” are absolute.
M101.6 Dimensional Tolerances. Dimensions are subject to
conventional industry tolerances for manufacturing processes,
material properties, and field conditions.
M101.7 Units of Measurement. Measurements are stated in
U.S. customary and metric units. The values stated in each system
(U.S. customary and metric units) may not be exact equivalents, and
each system shall be used independently of the other.
M102 Definitions
M102.1 Defined Terms. For the purpose of the MDE
Standards, the following terms have the indicated meaning:
End Transfer Surface. A transfer surface located at one
end of an examination surface that allows patient transfer at the
end and one adjoining side of the examination surface.
Enforcing Authority. An agency or other governmental
entity that adopts the MDE Standards as mandatory requirements for
entities subject to its jurisdiction. Enforcing authorities may
include, but are not limited to the United States Departments of
Justice and Health and Human Services.
Examination Chair. Diagnostic equipment with a seat in
which a patient typically is positioned with buttocks approximately
parallel to the ground and shins approximately perpendicular to the
ground. Examination chairs typically have back support and may
recline to properly position the patient during examination. Such
chairs may also have footrests or stirrups. Examination chairs
include, but are not limited to, equipment used for dental,
ophthalmic, podiatric, gynecological, urological, and ear, nose,
and throat examinations.
Imaging bed. A component of diagnostic scanning equipment
that accommodates patients in supine, prone, or side-lying
positions.
Imaging equipment with bores. Diagnostic scanning
equipment using magnets, x-rays, or detectors into which a patient
and the table on which the patient lies is inserted into the
equipment through a cylindrical opening (bore) in order to achieve
the positioning accuracy needed during the scan. Such equipment
includes, but is not limited to, computerized axial tomography (CT
or CAT), positron emission tomography (PET), and nuclear medicine
(NM) scanning equipment or a combination thereof.
Medical Diagnostic Equipment (Diagnostic Equipment).
Equipment used in, or in conjunction with, medical settings by
health care providers for diagnostic purposes.
Operable Parts. Components of diagnostic equipment that
are used by the patient to activate, deactivate, or adjust the
equipment.
Side Transfer Surface. A transfer surface located within
the length of the examination surface that allows patient transfer
on two opposing sides of the examination surface.
Transfer Surface. Part of diagnostic equipment onto which
patients who use mobility devices or aids transfer when moving onto
and off of the equipment.
Wheelchair Space. Space for a single wheelchair and its
occupant.
M102.2 Undefined Terms. Terms not defined in M102.1 or in
regulations or policies issued by an enforcing authority shall be
given their ordinarily accepted meaning in the sense that the
context implies.
M102.3 Interchangeability. Words, terms, and phrases used
in the singular include the plural and those used in the plural
include the singular.
Chapter 2: Scoping M201 General
M201.1 Application by Enforcing Authority. The enforcing
authority shall specify the number and type of diagnostic equipment
that are required to comply with the MDE Standards.
M201.2 General Exception. Medical diagnostic equipment
shall not be required to comply with one or more applicable
requirements in the MDE Standards in the rare circumstances where
compliance would alter diagnostically required structural or
operational characteristics of the equipment and would prevent the
use of the equipment for its intended diagnostic purpose.
Diagnostic equipment subject to M201.2 shall comply to the maximum
extent practicable.
Chapter 3: Technical Requirements M301 Diagnostic Equipment Used by
Patients in Supine, Prone, or Side-Lying Position
M301.1 General. Diagnostic equipment that supports
patients in a supine, prone, or side-lying position shall comply
with M301.
Exception: Examination chairs complying with M302 that
recline to facilitate diagnosis after patients transfer onto the
chair shall not be required to comply with M301.
M301.2 Transfer Surface. A transfer surface shall be
provided and shall comply with M301.2.
M301.2.1 Adjustability. Transfer surfaces shall be
adjustable in height measured from the floor to the top of the
uncompressed transfer surface and shall provide the following:
A. A low transfer position at a height of 17 inches (430 mm)
minimum and 19 inches (485 mm) maximum;
B. A high transfer position at 25 inches (635 mm); and
C. At least 4 additional transfer positions located between the
low and high transfer positions and separated by 1 inch (25 mm)
minimum.
M301.2.2 Sunset. The low transfer position height, Item A
of M301.2.1, shall cease to have effect on January 10, 2022.
M301.2.3 Size. The size of the transfer surface shall
comply with M301.2.3.1 or M301.2.3.2. The size of transfer surfaces
shall be measured from center points of their opposing sides.
M301.2.3.1 End Transfer Surface. End transfer surfaces
shall be 28 inches (710 mm) wide minimum and 17 inches (430 mm)
long minimum.
Exception: Transfer surfaces for imaging equipment with
bores shall be permitted to be 21 inches (535 mm) wide minimum but
shall not be permitted to be less than the full width of the
examination surface provided for the patient.
M301.2.3.2 Side Transfer Surface. Side transfer surfaces
shall be 28 inches (710 mm) wide minimum and 28 inches (710 mm)
long minimum.
Exception: Transfer surfaces for imaging equipment with
bores shall be permitted to be 21 inches (535 mm) wide minimum but
shall not be permitted to be less than the full width of the
examination surface provided for the patient.
M301.2.4 Unobstructed Transfer. Each transfer surface
shall provide two unobstructed sides for patient transfer.
Exceptions: 1. Obstructions no more than 3 inches (75 mm)
deep shall be permitted to extend beyond transfer sides of transfer
surfaces provided that such obstructions do not protrude above the
tops of transfer surfaces.
2. Temporary obstructions shall be permitted provided that they
can be repositioned during transfer to comply with M301.2.4,
including Exception 1.
M301.3 Supports. Transfer supports, leg supports, and
reclining surfaces shall comply with M301.3.
M301.3.1 Transfer Supports. Transfer surfaces required by
M301.2 shall provide transfer supports and shall comply with
M305.2.
M301.3.2 Leg Supports. Where stirrups are provided, leg
supports shall also be provided and shall comply with M305.4.
M301.3.3 Head and Back Support. Where the diagnostic
equipment is used in a reclined position, head and back support
shall be provided and shall comply with M305.5.
M301.4 Lift Compatibility. Diagnostic equipment shall be
usable with portable patient lifts and, when in use with such
lifts, shall comply with M301.4.1 or M301.4.2.
Exception: Where fixed overhead patient lifts are
provided, and when their use with diagnostic equipment is permitted
by an enforcing authority, diagnostic equipment shall not be
required to meet the lift compatibility requirements of this
section provided that such equipment is clearly labeled as not
compatible with portable floor lifts.
M301.4.1 Clearance in Base. The base of diagnostic
equipment shall provide a clearance 39 inches (990 mm) wide
minimum, 6 inches (150 mm) high minimum measured from the floor,
and 36 inches (915 mm) deep minimum measured from the edge of the
examination surface. Where the width of examination surfaces is
less than 36 inches (915 mm), the clearance depth shall extend the
full width of the equipment. Components of diagnostic equipment are
permitted to be located within 8 inches (205 mm) maximum of the
centerline of the clearance width.
M301.4.2 Clearance Around Base. The base of diagnostic
equipment shall provide a clearance 6 inches (150 mm) high minimum
measured from the floor and 36 inches (915 mm) deep minimum
measured from the edge of the examination surface. The width of the
base permitted within this clearance shall be 26 inches (660 mm)
wide maximum at the edge of the examination surface and shall be
permitted to increase at a rate of 1 inch (25 mm) in width for each
3 inches (75 mm) in depth.
M302 Diagnostic Equipment Used by Patients in Seated Position
M302.1 General. Diagnostic equipment that supports
patients in a seated position shall comply with M302.
Exception: Where weight scales contain wheelchair spaces
complying with M303 and also provide a seat integral to the
equipment, the scales shall not be required to comply with
M302.
M302.2 Transfer Surface. A transfer surface shall be
provided and shall comply with M302.2.
M302.2.1 Adjustability. Transfer surfaces shall be
adjustable in height measured from the floor to the top of the
uncompressed transfer surface and shall provide the following:
A. A low transfer position at a height of 17 inches (430 mm)
minimum and 19 inches (485 mm) maximum;
B. A high transfer position at 25 inches (635 mm); and
C. At least 4 additional transfer positions located between the
low and high transfer positions and separated by 1 inch (25 mm)
minimum.
M302.2.2 Sunset. The low transfer position height, Item A
of M302.2.1, shall cease to have effect on January 10, 2022.
M302.2.3 Size. Transfer surfaces shall be 21 inches (610
mm) wide minimum and 17 inches (430 mm) deep minimum. The size of
transfer surfaces shall be measured from center points of their
opposing sides.
M302.2.4 Transfer Sides. Options to transfer from a
mobility device shall be provided on two adjoining sides of
transfer surfaces.
Exception: Options to transfer to or from a mobility
device onto opposing sides of transfer surfaces shall be permitted
where the transfer surface is obstructed by fixed footrests.
M302.2.5 Unobstructed Transfer. Each transfer side
complying with M302.2.4 shall provide unobstructed access to
transfer surfaces.
Exceptions: 1. Obstructions no more than 3 inches (75 mm)
deep shall be permitted to extend beyond transfer sides of transfer
surfaces provided that such obstructions do not protrude above the
tops of transfer surfaces.
2. Temporary obstructions shall be permitted provided that they
can be repositioned during transfer to comply with M302.2.5,
including Exception 1.
M302.3 Supports. Transfer supports, leg supports and
reclining surfaces shall comply with M302.3.
M302.3.1 Transfer Supports. Transfer supports shall be
provided for use with transfer sides required by M302.2.4 and shall
comply with M305.2.1.1, M305.2.2.1, and M305.2.3 through
M305.2.8.
M302.3.2 Leg Supports. Where stirrups are provided, leg
supports shall also be provided and comply with M305.4.
M302.3.3 Head and Back Support. Where the diagnostic
equipment is used in a reclined position, head and back support
shall be provided and shall comply with M305.5.
M302.4 Lift Compatibility. Diagnostic equipment shall be
usable with portable patient lifts and, when in use with such
lifts, shall comply with M302.4.1 or M302.4.2.
Exception: Where fixed overhead patient lifts are
provided, and when their use with diagnostic equipment is permitted
by an enforcing authority, diagnostic equipment shall not be
required to meet the lift compatibility requirements of this
section provided that such equipment is clearly labeled as not
compatible with portable floor lifts.
M302.4.1 Clearance in Base. The base of the diagnostic
equipment shall provide a clearance 39 inches (990 mm) wide
minimum, 6 inches (150 mm) high minimum measured from the floor,
and 36 inches (915 mm) deep minimum measured from the edge of the
examination surface. Where the width of the examination surface is
less than 36 inches (915 mm), the clearance depth shall extend the
full width of the equipment. Equipment components are permitted to
be located within 8 inches (205 mm) maximum of the centerline of
the clearance width.
M302.4.2 Clearance Around Base. The base of the
diagnostic equipment shall provide a clearance 6 inches (150 mm)
high minimum measured from the floor and 36 inches (915 mm) deep
minimum measured from the edge of the examination surface. The
width of the base permitted within this clearance shall be 26
inches (660 mm) wide maximum at the edge of the examination surface
and shall be permitted to increase at a rate of 1 inch (25 mm) in
width for each 3 inches (75 mm) in depth.
M303 Diagnostic Equipment Used by Patients Seated in a Wheelchair
M303.1 General. Diagnostic equipment used by patients
seated in a wheelchair shall comply with M303.
M303.2 Wheelchair Spaces. Wheelchair spaces complying
with M303.2 shall be provided at diagnostic equipment.
M303.2.1 Orientation. Wheelchair spaces shall be designed
so that a patient seated in a wheelchair orients in the same
direction that a patient not seated in a wheelchair orients when
the diagnostic equipment is in use.
M303.2.2 Width. Wheelchair spaces shall be 36 inches (915
mm) wide minimum.
Exception: Wheelchair spaces located on raised platforms
shall be permitted to be 32 inches (815 mm) wide minimum to a
height of 4 inches (100 mm) measured from the platform surface.
M303.2.3 Depth. The depth of wheelchair spaces shall
comply with M303.2.3.
M303.2.3.1 Front or Rear Entry. Where wheelchair space
entry and exit is provided at only one end (front or rear) the
wheelchair space shall be 48 inches (1220 mm) deep minimum.
M303.2.3.2 Pass Through Entry. Where wheelchair space
entry and exit permits pass through from one end to the other, the
wheelchair space shall be 40 inches deep (1015 mm) minimum.
M303.2.3.3 Side Entry. Where wheelchair space entry is
only from the side, the wheelchair space shall be 60 inches (1525
mm) deep minimum.
M303.2.4 Equipment Clearances. Where wheelchair spaces
are entered from the rear and includes space beneath components,
wheelchair spaces shall include knee and toe clearances complying
with M303.2.4.1 for breast platforms and M303.2.4.2 for all other
equipment.
M303.2.4.1 Breast Platforms. Wheelchair spaces beneath
breast platforms shall comply with M303.2.4.1.
M303.2.4.1.1 Depth. Wheelchair spaces shall include knee
and toe clearance 25 inches (635 mm) deep minimum and 28 inches
(710 mm) deep maximum.
M303.2.4.1.2 Height. Wheelchair spaces shall include toe
clearance 9 inches (230 mm) high minimum above the floor measured
to a depth of 6 inches (150 mm) maximum from the toe end of the
wheelchair space. Knee clearance shall be provided at a depth of 19
inches (485 mm) minimum and 22 inches (560 mm) maximum at 9 inches
(230 mm) above the floor and at a depth of 16 inches (405 mm)
minimum at 27 inches (685 mm) above the floor measured from the
leading edge of the breast platform. Between 9 inches (230 mm) and
27 inches (685 mm) above the floor, the knee clearance shall be
permitted to reduce at a rate of 1 inch (25 mm) in depth for every
6 inches (150 mm) in height.
Exception: Components shall be permitted to extend into
the wheelchair space at a height of 1 1/2 inches (38 mm) maximum
between 17 inches (430 mm) minimum and 25 inches (635 mm) maximum
in depth measured from the leading edge of the breast platform.
From 25 inches (635 mm) to 28 inches (710 mm) in depth the height
of a component above 1 1/2 inches (38 mm) shall be beveled at a
rate of 2.5:3 maximum.
M303.2.4.2 Other Equipment. Wheelchair spaces beneath
diagnostic equipment other than breast platforms shall comply with
M303.2.4.2.
M303.2.4.2.1 Depth. Wheelchair spaces shall include knee
and toe clearance 17 inches (430 mm) deep minimum and 25 inches
(635 mm) deep maximum.
M303.2.4.2.2 Height. Wheelchair spaces shall include toe
clearance 9 inches (230 mm) high minimum above the floor measured
to a depth of 6 inches (150 mm) maximum measured from the toe end
of the wheelchair space. Knee clearance shall be provided at a
depth of 11 inches (280 mm) minimum and 25 inches (635 mm) maximum
at 9 inches (230 mm) above the floor and at a depth of 8 inches
(205 mm) minimum at 27 inches (685 mm) above the floor measured
from the leading edge of the equipment. Between 9 inches (230 mm)
and 27 inches (685 mm) above the floor, the knee clearance shall be
permitted to reduce at a rate of 1 inch (25 mm) in depth for every
6 inches (150 mm) in height.
M303.2.5 Surfaces. Wheelchair space surfaces shall not
slope more than 1:48 in any direction.
M303.2.6 Edge Protection. Where wheelchair spaces are
provided on a platform raised more than 1 1/2 inches (38 mm) in
height, edge protection 2 inches (51 mm) high minimum measured from
the surface of the platform shall be provided on each side not
providing entry to or exit from the equipment.
M303.3 Entry. Where there is a change in level at the
entry to wheelchair spaces, the change in level shall comply with
M303.3.
M303.3.1 Vertical. Changes in level of 1/4 inch (6.4 mm)
high maximum shall be permitted to be vertical.
M303.3.2 Beveled. Changes in level between 1/4 inch (6.4
mm) high and 1/2 inch (13 mm) high maximum shall be beveled with a
slope not steeper than 1:2.
M303.3.3 Ramped. Changes in level greater than 1/2 inch
(13 mm) high shall be ramped and shall comply with M303.3.3.
M303.3.3.1 Running Slope. Ramp runs shall have a running
slope not steeper than 1:12.
Exception: A running slope not steeper than 1:8 shall be
permitted for ramp runs with a maximum height of 2 1/2 inches (64
mm).
M303.3.3.2 Cross Slope. The cross slope of ramp runs
shall not be steeper than 1:48.
M303.3.3.3 Clear Width. The clear width of ramp runs
shall be 36 inches (915 mm) minimum.
M303.3.3.4 Edge Protection. Ramps with drop offs 1/2 inch
(13 mm) or greater shall provide edge protection 2 inches (50 mm)
high minimum on each side with a drop off.
M303.3.3.5 Handrails. Ramps with a rise greater than 6
inches (150 mm) shall provide handrails on both sides.
M303.4 Components. Where components of diagnostic
equipment are used to examine specific body parts, the components
shall be capable of examining the body parts of a patient seated in
a wheelchair. Breast platforms shall comply with M303.4.1.
M303.4.1 Breast Platform Adjustability. Breast platforms
shall be continuously adjustable from a low height of 26 inches
(660 mm) to a high height of 42 inches (1065 mm) above the
floor.
M304 Diagnostic Equipment Used by Patients in Standing Position
M304.1 General. Diagnostic equipment used by patients in
a standing position shall comply with M304.
M304.2 Standing Surface. Equipment surfaces on which
patients stand must comply with M304.2
M304.2.1 Slip Resistant. The surface on which the patient
stands shall be slip resistant.
M304.2.2 Standing Supports. Standing supports shall be
provided on two sides of the standing surface and shall comply with
M305.3.
Exception: Diagnostic equipment with entry and exit
permitting pass-through from one end to the other shall be
permitted to provide one standing support on one side of the
standing surface provided that the standing support complies with
the requirements for standing supports in a horizontal position in
M305.3.
M305 Supports
M305.1 General. Supports shall comply with M305.
M305.2 Transfer Supports. Transfer supports shall comply
with M305.2.
M305.2.1 Location. Transfer supports shall comply with
M305.2.1.1 or M305.2.1.2 and shall be located 1 1/2 inches (38 mm)
maximum measured horizontally from the plane defined by the nearest
edge of the transfer surface.
Exception: Where the support folds, collapses, or
articulates, the transfer support shall be permitted to be located
3 inches (75 mm) maximum from the plane defined by the nearest edge
of the transfer surface.
M305.2.1.1 End Transfer Supports. Transfer supports for
transfer surfaces complying with M301.2.3.1 and M302.2 shall be
located on the short side (length) opposite the transfer side.
M305.2.1.2 Side Transfer Supports. Transfer supports for
transfer surfaces complying with M301.2.3.2 shall be capable of
supporting transfer on each side of the transfer surface.
M305.2.2 Length. The length of transfer supports shall
comply with M305.2.2.1 or M305.2.2.2.
M305.2.2.1 End Transfer Supports. Transfer supports for
transfer surfaces complying with M301.2.3.1 and M305.2.2.1 shall be
15 inches (380 mm) long minimum. Transfer supports shall be
positioned along 13 1/2 inches (345 mm) minimum of the depth of the
transfer surface.
M305.2.2.2 Side Transfer Supports. Transfer supports for
transfer surfaces complying with M301.2.3.2 shall be 28 inches (710
mm) long minimum and shall be positioned along the width of
transfer surfaces.
Exceptions: 1. Where transfer surfaces are part of an
articulating surface, the support shall be permitted to be 15
inches (380 mm) long minimum.
2. Where the width of an imaging bed is more than 24 inches (533
mm), transfer supports shall be permitted to be 12 inches (305 mm)
long minimum.
M305.2.3 Height. During use, the tops of transfer support
gripping surfaces shall be 6 inches (150 mm) minimum and 19 inches
(485 mm) maximum higher than the top of the associated uncompressed
transfer surface.
Exception: Where the width of the transfer surface for
imaging beds exceed 24 inches (610 mm), the tops of the gripping
surfaces shall be permitted to be 3 inches (75 mm) minimum and 6
inches (150 mm) maximum higher than the top of the associated
uncompressed transfer surface.
M305.2.4 Cross Section. Transfer supports shall have a
cross section complying with 305.2.4.1 or 305.2.4.2.
M305.2.4.1 Circular Cross Section. Transfer supports with
circular cross sections shall have an outside diameter of 1 1/4
inches (32 mm) minimum and 2 inches (51 mm) maximum.
M305.2.4.2 Non-Circular Cross Section. Transfer supports
with non-circular cross sections shall have a cross-section
dimension of 2 inches (51 mm) maximum and a perimeter dimension of
4 inches (100 mm) minimum and 4.8 inches (120 mm) maximum.
M305.2.5 Surface Hazards. Transfer supports and surfaces
adjacent to transfer supports shall be free of sharp or abrasive
components and shall have eased edges.
M305.2.6 Gripping Surface. Transfer support gripping
surfaces shall be continuous along their length and shall not be
obstructed along their tops or sides. The bottoms of transfer
support gripping surfaces shall not be obstructed for more than 20
percent of their length.
M305.2.7 Clearance. Clearance between the transfer
support gripping surface and adjacent surfaces or obstructions
shall be 1 1/2 inches (38 mm) minimum.
M305.2.8 Fittings. Transfer supports shall not rotate
within their fittings when in place for transfer.
M305.3 Standing Supports. Standing supports shall provide
continuous support throughout use of the diagnostic equipment and
shall comply with M305.3.
M305.3.1 Length. The length of gripping surfaces for
standing supports shall be based on the position of the standing
supports in relation to the standing surfaces they serve.
Horizontal standing support gripping surfaces shall comply with
M305.3.1.1, horizontal standing support gripping surfaces on
diagnostic equipment containing a wheelchair space shall comply
with M305.3.1.2 and, vertical standing support gripping surfaces
shall comply with M305.3.1.3.
M305.3.1.1 Horizontal Position. The length of gripping
surfaces on horizontal standing supports shall be 4 inches (100 mm)
minimum except for diagnostic equipment containing a wheelchair
space which shall comply with M305.3.1.2.
M305.3.1.2 Diagnostic Equipment Containing a Wheelchair
Space. On diagnostic equipment containing wheelchair spaces
with one entry that also serves as the exit, the length of the
gripping surface of horizontal standing supports shall be equal to
or greater than 80 percent of the overall length of the platform.
On diagnostic equipment containing a wheelchair space and
permitting pass-through from one end to the other, the length of
the gripping surface on horizontal standing supports shall be at
least equal to the length of the platform.
M305.3.1.3 Vertical Position. The length of the gripping
surface on vertical standing supports shall be 18 inches (455 mm)
minimum.
M305.3.2 Height. The height of gripping surfaces for
standing supports shall be based on the position of the standing
supports in relation to the standing surfaces they serve.
Horizontal standing support gripping surfaces shall comply with
M305.3.2.1 and vertical standing support gripping surfaces shall
comply with M305.3.2.2.
M305.3.2.1 Horizontal Position. The height of the top of
the gripping surface on horizontal standing supports shall be 34
inches (865 mm) minimum and 38 inches (965 mm) maximum above the
standing surface.
M305.3.2.2 Vertical Position. The height of the lowest
end of the gripping surface on vertical standing supports shall be
34 inches (865 mm) minimum and 37 inches (940 mm) maximum above the
standing surface.
M305.3.3 Fittings. Standing supports shall not rotate
within their fittings.
M305.4 Leg Supports. Leg supports shall provide a method
of supporting, positioning, and securing the patient's legs.
M305.5 Head and Back Support. Where the diagnostic
equipment is used in a reclined position, head and back support
shall be provided. Where the incline of the back support can be
modified while in use, head and back support shall be provided
throughout the entire range of the incline.
M306 Communication
M306.1 General. Where instructions or other information
necessary for performance of the diagnostic procedure is
communicated to the patient through the diagnostic equipment, the
instructions and other information shall be provided in at least
two of the following methods: Audible, visible, or tactile.
M307 Operable Parts
M307.1 General. Operable parts for patient use shall
comply with M307.
M307.2 Tactilely Discernible. Operable parts shall be
tactilely discernible without activation.
M307.3 Operation. Operable parts shall be operable with
one hand and shall not require tight grasping, pinching, or
twisting of the wrist.
M307.4 Operating Force. The force required to activate
operable parts shall be 5 pounds (22.2 N) maximum.