Appendix A to Part 718 - Standards for Administration and Interpretation of Chest Radiographs (X-rays)
20:4.0.2.2.6.4.74.7.1 : Appendix A
Appendix A to Part 718 - Standards for Administration and
Interpretation of Chest Radiographs (X-rays)
The following standards are established in accordance with
sections 402(f)(1)(D) and 413(b) of the Act. They were developed in
consultation with the National Institute for Occupational Safety
and Health (NIOSH) of the Centers for Disease Control and
Prevention in the Department of Health and Human Services. These
standards are promulgated for the guidance of physicians and
medical technicians to ensure that uniform procedures are used in
administering and interpreting X-rays and that the best available
medical evidence will be submitted in connection with a claim for
black lung benefits. If it is established that one or more
standards have not been met, the claims adjudicator may consider
such fact in determining the evidentiary weight to be assigned to
the physician's report of an X-ray.
(a) Definitions. (1) Digital radiography systems,
as used in this context, include both digital radiography (DR) and
computed radiography (CR). Digital radiography is the term used for
digital X-ray image acquisition systems in which the X-ray signals
received by the image detector are converted nearly instantaneously
to electronic signals without moveable cassettes. Computed
radiography is the term for digital X-ray image acquisition systems
that detect X-ray signals using a cassette-based photostimulable
storage phosphor. Subsequently, the cassette is processed using a
stimulating laser beam to convert the latent radiographic image to
electronic signals which are then processed and stored so they can
be displayed.
(2) Qualified medical physicist means an individual who
is trained in evaluating the performance of radiographic equipment
including radiation controls and facility quality assurance
programs, and has the relevant current certification by a competent
U.S. national board, or unrestricted license or approval from a
U.S. State or Territory.
(3) Radiographic technique chart means a table that
specifies the types of cassette, intensifying screen, film or
digital detector, grid, filter, and lists X-ray machine settings
(timing, kVp, mA) that enables the radiographer to select the
correct settings based on the body habitus or the thickness of the
chest tissue.
(4) Radiologic technologist means an individual who has
met the requirements for privileges to perform general radiographic
procedures and for competence in using the equipment and software
employed by the examining facility to obtain chest images as
specified by the State or Territory and examining facility in which
such services are provided. Optimally, such an individual will have
completed a formal training program in radiography leading to a
certificate, an associate's degree, or a bachelor's degree and
participated in the voluntary initial certification and annual
renewal of registration for radiologic technologists offered by the
American Registry of Radiologic Technologists.
(5) Soft copy means the image of a coal miner's chest
radiograph acquired using a digital radiography system, viewed at
the full resolution of the image acquisition system using an
electronic medical image display device.
(b) General provisions. (1) Facilities must maintain
ongoing licensure and certification under relevant local, State,
and Federal laws and regulations for all digital equipment and
related processes covered by this Appendix. Radiographic equipment,
its use and the facilities (including mobile facilities) in which
such equipment is used must conform to applicable State or
Territorial and Federal regulations. Where no applicable
regulations exist regarding reducing the risk from ionizing
radiation exposure in the clinical setting, radiographic equipment,
its use and the facilities (including mobile facilities) in which
such equipment is used should conform to the recommendations in
NCRP Report No. 102, NCRP Report No. 105, and NCRP Report No. 147
(incorporated by reference, see § 718.5).
(2) Chest radiographs of miners must be performed:
(i) By or under the supervision of a physician who makes chest
radiographs in the normal course of practice and who has
demonstrated ability to make chest radiographs of a quality to best
ascertain the presence of pneumoconiosis; or
(ii) By a radiologic technologist.
(3) Miners must be disrobed from the waist up at the time the
radiograph is given. The facility must provide a dressing area and
for those miners who wish to use one, the facility will provide a
clean gown. Facilities must be heated to a comfortable
temperature.
(4) Before the miner is advised that the examination is
concluded, the radiograph must be processed and inspected and
accepted for quality standards by the physician, or if the
physician is not available, acceptance may be made by the
radiologic technologist. In a case of a substandard radiograph,
another must be made immediately.
(c) Chest radiograph specifications - film. (1) Every
chest radiograph must be a single posteroanterior projection at
full inspiration on a film being no less than 14 by 17 inch film.
Additional chest films or views must be obtained if they are
necessary for clarification and classification. The film and
cassette must be capable of being positioned both vertically and
horizontally so that the chest radiograph will include both apices
and costophrenic angles. If a miner is too large to permit the
above requirements, then a projection with minimum loss of
costophrenic angle must be made.
(2) Radiographs must be made with a diagnostic X-ray machine
having a rotating anode tube with a maximum of a 2 mm source (focal
spot).
(3) Except as provided in paragraph (c)(4) of this appendix,
radiographs must be made with units having generators that comply
with the following:
(i) Generators of existing radiographic units acquired by the
examining facility prior to July 27, 1973, must have a minimum
rating of 200 mA at 100 kVp;
(ii) Generators of units acquired subsequent to that date must
have a minimum rating of 300 mA at 125 kVp. A generator with a
rating of 150 kVp is recommended.
(4) Radiographs made with battery-powered mobile or portable
equipment must be made with units having a minimum rating of 100 mA
at 110 kVp at 500 Hz, or 200 mA at 110 kVp at 60 Hz.
(5) Capacitor discharge and field emission units may be
used.
(6) Radiographs must be given only with equipment having a
beam-limiting device that does not cause large unexposed
boundaries. The use of such a device must be discernible from an
examination of the radiograph.
(7) To ensure high quality chest radiographs:
(i) The maximum exposure time must not exceed 50 milliseconds
except that with single phase units with a rating less than 300 mA
at 125 kVp and subjects with chests over 28 cm postero-anterior,
the exposure may be increased to not more than 100
milliseconds;
(ii) The source or focal spot to film distance must be at least
6 feet.
(iii) Medium-speed film and medium-speed intensifying screens
are recommended. However, any film-screen combination, the rated
“speed” of which is at least 100 and does not exceed 300, which
produces radiographs with spatial resolution, contrast, latitude
and quantum mottle similar to those of systems designated as
“medium speed” may be employed;
(iv) Film-screen contact must be maintained and verified at
6-month or shorter intervals.
(v) Intensifying screens must be inspected at least once a month
and cleaned when necessary by the method recommended by the
manufacturer;
(vi) All intensifying screens in a cassette must be of the same
type and made by the same manufacturer;
(vii) When using over 90 kV, a suitable grid or other means of
reducing scattered radiation must be used;
(viii) The geometry of the radiographic system must ensure that
the central axis (ray) of the primary beam is perpendicular to the
plane of the film surface and impinges on the center of the
film.
(8) Radiographic processing:
(i) Either automatic or manual film processing is acceptable. A
constant time-temperature technique must be meticulously employed
for manual processing.
(ii) If mineral or other impurities in the processing water
introduce difficulty in obtaining a high-quality radiograph, a
suitable filter or purification system must be used.
(9) An electric power supply must be used that complies with the
voltage, current, and regulation specified by the manufacturer of
the machine.
(10) A test object may be required on each radiograph for an
objective evaluation of film quality at the discretion of the
Department of Labor.
(11) Each radiograph made under this Appendix must be
permanently and legibly marked with the name and address of the
facility at which it is made, the miner's DOL claim number, the
date of the radiograph, and left and right side of the film. No
other identifying markings may be recorded on the radiograph.
(d) Chest radiograph specifications - digital radiography
systems. (1) Every digital chest radiograph must be a single
posteroanterior projection at full inspiration on a digital
detector with sensor area being no less than 1505 square
centimeters with a minimum width of 35 cm. The imaging plate must
have a maximum pixel pitch of 200 µm, with a minimum bit depth of
10. Spatial resolution must be at least 2.5 line pairs per
millimeter. The storage phosphor cassette or digital image detector
must be positioned either vertically or horizontally so that the
image includes the apices and costophrenic angles of both right and
left lungs. If the detector cannot include the apices and
costophrenic angles of both lungs as described, then the two
side-by-side images can be obtained that together include the
apices and costophrenic angles of both right and left lungs.
(2) Radiographs must be made with a diagnostic X-ray machine
with a maximum actual (not nominal) source (focal spot) of 2 mm, as
measured in two orthogonal directions.
(3) Radiographs must be made with units having generators which
have a minimum rating of 300 mA at 125 kVp. Exposure kilovoltage
must be at least the minimum as recommended by the manufacturer for
chest radiography.
(4) An electric power supply must be used that complies with the
voltage, current, and regulation specified by the manufacturer of
the machine. If the manufacturer or installer of the radiographic
equipment recommends equipment for control of electrical power
fluctuations, such equipment must be used as recommended.
(5) Radiographs must be obtained only with equipment having a
beam-limiting device that does not cause large unexposed
boundaries. The beam limiting device must provide rectangular
collimation. Electronic post-image acquisition “shutters” available
on some CR or DR systems that limit the size of the final image and
that simulate collimator limits must not be used. The use and
effect of the beam limiting device must be discernible on the
resulting image.
(6) Radiographic technique charts must be used that are
developed specifically for the X-ray system and detector
combinations used, indicating exposure parameters by anatomic
measurements.
(7) To ensure high quality chest radiographs:
(i) The maximum exposure time must not exceed 50 milliseconds
except for subjects with chests over 28 cm posteroanterior, for
whom the exposure time must not exceed 100 milliseconds.
(ii) The distance from source or focal spot to detector must be
at least 70 inches (or 180 centimeters if measured in
centimeters).
(iii) The exposure setting for chest images must be within the
range of 100-300 equivalent exposure speeds and must comply with
ACR Practice Guidelines for Diagnostic Reference Levels in Medical
X-ray Imaging, Section V - Diagnostic Reference Levels for Imaging
with Ionizing Radiation and Section VII-Radiation Safety in Imaging
(incorporated by reference, see § 718.5). Radiation
exposures should be periodically measured and patient radiation
doses estimated by the medical physicist to assure doses are as low
as reasonably achievable.
(iv) Digital radiography system performance, including
resolution, modulation transfer function (MTF), image
signal-to-noise and detective quantum efficiency must be evaluated
and judged acceptable by a qualified medical physicist using the
specifications in AAPM Report No. 93, pages 1-68 (incorporated by
reference, see § 718.5). Image management software and
settings for routine chest imaging must be used, including routine
amplification of digital detector signal as well as standard image
post-processing functions. Image or edge enhancement software
functions must not be employed unless they are integral to the
digital radiography system (not elective); in such cases, only the
minimum image enhancement permitted by the system may be
employed.
(v)(A) The image object, transmission and associated data
storage, film format, and transmissions of associated information
must conform to the following components of the Digital Imaging and
Communications in Medicine (DICOM) standard (incorporated by
reference, see § 718.5):
(1) DICOM Standard PS 3.3-2011, Annex A - Composite
Information Object Definitions, sections: Computed Radiographic
Image Information Object Definition; Digital X-Ray Image
Information Object Definition; X-Ray Radiation Dose SR Information
Object Definition; and Grayscale Softcopy Presentation State
Information Object Definition.
(2) DICOM Standard PS 3.4-2011: Annex B - Storage Service
Class; Annex N - Softcopy Presentation State Storage SOP Classes;
Annex O - Structured Reporting Storage SOP Classes.
(3) DICOM Standard PS 3.10-2011.
(4) DICOM Standard PS 3.11-2011.
(5) DICOM Standard PS 3.12-2011.
(6) DICOM Standard PS 13.14-2011.
(7) DICOM Standard PS 3.16-2011.
(B) Identification of each miner, chest image, facility, date
and time of the examination must be encoded within the image
information object, according to DICOM Standard PS 3.3-2011,
Information Object Definitions, for the DICOM “DX” object. If data
compression is performed, it must be lossless. Exposure parameters
(kVp, mA, time, beam filtration, scatter reduction, radiation
exposure) must be stored in the DX information object.
(C) Exposure parameters as defined in the DICOM Standard PS
3.16-2011 must additionally be provided when such parameters are
available from the facility digital image acquisition system or
recorded in a written report or electronic file and transmitted to
OWCP.
(8) A specific test object may be required on each radiograph
for an objective evaluation of image quality at the Department of
Labor's discretion.
(9) CR imaging plates must be inspected at least once a month
and cleaned when necessary by the method recommended by the
manufacturer.
(10) A grid or air gap for reducing scattered radiation must be
used; grids must not be used that cause Moiré interference patterns
in either horizontal or vertical images.
(11) The geometry of the radiographic system must ensure that
the central axis (ray) of the primary beam is perpendicular to the
plane of the CR imaging plate or DR detector and is correctly
aligned to the grid.
(12) Radiographs must not be made when the environmental
temperatures and humidity in the facility are outside the
manufacturer's recommended range of the CR and DR equipment to be
used.
(13) All interpreters, whenever classifying digitally acquired
chest radiographs, must have immediately available for reference a
complete set of ILO standard digital chest radiographic images
provided for use with the Guidelines for the Use of the ILO
International Classification of Radiographs of Pneumoconioses (2011
Revision) (incorporated by reference, see § 718.5).
Modification of the appearance of the standard images using
software tools is not permitted.
(14) Viewing systems should enable readers to display the coal
miner's chest image at the full resolution of the image acquisition
system, side-by-side with the selected ILO standard images for
comparison.
(i)(A) Image display devices must be flat panel monitors
displaying at least 3 MP at 10 bit depth. Image displays and
associated graphics cards must meet the calibration and other
specifications of the Digital Imaging and Communications in
Medicine (DICOM) standard PS 3.14-2011 (incorporated by reference,
see § 718.5).
(B) Image displays and associated graphics cards must not
deviate by more than 10 percent from the grayscale standard display
function (GSDF) when assessed according to the AAPM On-Line Report
No. 03, pages 1-146 (incorporated by reference, see §
718.5).
(ii) Display system luminance (maximum and ratio), relative
noise, linearity, modulation transfer function (MTF), frequency,
and glare should meet or exceed recommendations listed in AAPM
On-Line Report No. 03, pages 1-146 (incorporated by reference,
see § 718.5). Viewing displays must have a maximum luminance
of at least 171 cd/m 2, a ratio of maximum luminance to minimum
luminance of at least 250, and a glare ratio greater than 400. The
contribution of ambient light reflected from the display surface,
after light sources have been minimized, must be included in
luminance measurements.
(iii) Displays must be situated so as to minimize front surface
glare. Readers must minimize reflected light from ambient sources
during the performance of classifications.
(iv) Measurements of the width and length of pleural shadows and
the diameter of opacities must be taken using calibrated software
measuring tools. If permitted by the viewing software, a record
must be made of the presentation state(s), including any noise
reduction and edge enhancement or restoration functions that were
used in performing the classification, including any annotations
and measurements.
(15) Quality control procedures for devices used to display
chest images for classification must comply with the
recommendations of the American Association of Physicists in
Medicine AAPM On-Line Report No. 03, pages 1-146 (incorporated by
reference, see § 718.5). If automatic quality assurance
systems are used, visual inspection must be performed using one or
more test patterns recommended by the medical physicist every 6
months, or more frequently, to check for defects that automatic
systems may not detect.
(16) Classification of CR and DR digitally-acquired chest
radiographs under this Part must be performed based on the viewing
images displayed as soft copies using the viewing workstations
specified in this section. Classification of radiographs must not
be based on the viewing of hard copy printed transparencies of
images that were digitally-acquired.
(17) The classification of chest radiographs based on digitized
copies of chest radiographs that were originally acquired using
film-screen techniques is not permissible.
[79 FR 21613, Apr. 17, 2014]